Chapter 3. Operation and Programming
Moisture Target Series 6 Hygrometer User’s Manual
45
3.4.4 Selecting an Alarm Type (cont.)
•
SetPoint
: Alarm activates when parameter exceeds upper limit, and
deactivates when parameter is less than lower limit.
•
Inner Band
: Alarm activates when parameter is between upper and
lower limits.
•
Outer Band
: Alarm activates when parameter is outside upper and
lower limits.
3.4.5 How the Alarm Types Work
Figure 45: Example of Alarm Types