Chapter 5. Specifications
Moisture Target Series 6 User’s Manual
85
Chapter 5. Specifications
5.1 Electronics
Input:
moisture signal from GE thin-film aluminum oxide moisture sensor on
an
M Series
probe or a
VeriDri
probe
Intrinsic Safety:
external safety barrier for moisture input (optional)
Analog Output:
single, isolated recorder output for dew point,
internally optically isolated, 10-bit (0.1%) resolution
0–2 V: 10 k
minimum load resistance
0–20 mA: 400
maximum series resistance
4–20 mA: 400
maximum series resistance
Outputs are user-programmable within the range of the instrument and
the corresponding probe.
Alarm Relays:
1 fault alarm and 2 programmable high/low alarms:
Standard designs are available for the high/low alarms, set to trip at any
level within the range of the instrument, and programmable from the
front panel.
Form C SPDT Relays:
Standard
3A @ 250VAC
3A @ 30VDC
Содержание Moisture Target Series 6
Страница 2: ......
Страница 4: ...ii no content intended for this page ...
Страница 10: ...Contents viii Moisture Target Series 6 Hygrometer User s Manual ...
Страница 90: ...Chapter 3 Operation and Programming 80 Moisture Target Series 6 Hygrometer User s Manual ...
Страница 103: ...Appendix B Menu Maps Moisture Target Series 6 Hygrometer User s Manual 93 Appendix B Menu Maps ...
Страница 104: ...Appendix B Menu Maps 94 Moisture Target Series 6 Hygrometer User s Manual Figure 49 Main Menu Map Using M Series Probe ...
Страница 105: ...Appendix B Menu Maps Moisture Target Series 6 Hygrometer User s Manual 95 Figure 50 Main Menu Map Using VeriDri Probe ...
Страница 106: ...Appendix B Menu Maps 96 Moisture Target Series 6 Hygrometer User s Manual no content intended for this page ...
Страница 126: ...Index 116 Moisture Target Series 6 User s Manual ...
Страница 130: ......
Страница 131: ......