Chapter 3. Operation and Programming
66
Moisture Target Series 6 Hygrometer User’s Manual
3.6.6a Setting the Calibration High Reference
To update high reference settings, from the
Edit Cal Refs
menu select
High Reference
and press
Enter
. A screen similar to the
following appears.
Use the left and right arrow keys to select
each digit to be changed and the up and
down arrow keys to increase or decrease its
value. Once the changes have been made,
press
Enter
. The screen returns to the
Edit Cal Refs
menu.
3.6.6b Setting the Calibration Low Reference
To update low reference settings, from the
Edit Cal Refs
menu select
Low Reference
and press
Enter
. A screen similar to the
following appears.
Use the left and right arrow keys to select
each digit to be changed and the up and
down arrow keys to increase or decrease its
value. Once the changes have been made,
press
Enter
. The screen returns to the
Edit Cal Refs
menu.
Содержание Moisture Target Series 6
Страница 2: ......
Страница 4: ...ii no content intended for this page ...
Страница 10: ...Contents viii Moisture Target Series 6 Hygrometer User s Manual ...
Страница 90: ...Chapter 3 Operation and Programming 80 Moisture Target Series 6 Hygrometer User s Manual ...
Страница 103: ...Appendix B Menu Maps Moisture Target Series 6 Hygrometer User s Manual 93 Appendix B Menu Maps ...
Страница 104: ...Appendix B Menu Maps 94 Moisture Target Series 6 Hygrometer User s Manual Figure 49 Main Menu Map Using M Series Probe ...
Страница 105: ...Appendix B Menu Maps Moisture Target Series 6 Hygrometer User s Manual 95 Figure 50 Main Menu Map Using VeriDri Probe ...
Страница 106: ...Appendix B Menu Maps 96 Moisture Target Series 6 Hygrometer User s Manual no content intended for this page ...
Страница 126: ...Index 116 Moisture Target Series 6 User s Manual ...
Страница 130: ......
Страница 131: ......