moisture.IQ User’s Manual
55
Chapter 4. Using the Outputs, Alarms and Logger Menus
4.2.3 Setting the Alarm Range Error Response
A
Range Error
can occur when a measured value is within the capacity of the analyzer, but it exceeds the
calibration range of the probe. They can be either
Under Range
or
Over Range
errors.
The moisture.IQ indicates
Range Errors
with an
Over Range
or
Under Range
message.The error condition
applies to all measurements of the same mode. For example, if a dew point measurement is
Over Range
, then
moisture in ppMv will also be
Over Range.
If multiple
Range Errors
occur simultaneously, the moisture.IQ
responds to them in the following order:
Oxygen Errors
,
Moisture Errors
,
Temperature Errors
and then
Pressure
Errors.
Complete the following steps to configure the range error response of the selected alarm:
1.
Tap the
Channel (1-6)
and
Alarm
(
A
or
B
), and then tap the
Range Error
button to open a screen similar
below.
Figure 49: Range Error Setup Screen - Alarm Response
2.
Set the alarm to either
Trip
on
Under Range
errors or to
Ignore
Under Range
errors (
Ignore
is the default
setting).
3.
Set the alarm to either
Trip
on
Over Range
errors or to
Ignore
Over Range
errors (
Ignore
is the default
setting).
4.
Tap the
Save
button to keep the new settings or tap the
Cancel
button to keep the previous settings.
Содержание moisture.IQ
Страница 1: ...Moisture and Oxygen 910 298 Rev E September 2017 bhge com moisture IQ User s Manual ...
Страница 2: ......
Страница 4: ... no content intended for this page ...
Страница 12: ...Preface x moisture IQ User s Manual no content intended for this page ...
Страница 45: ...moisture IQ User s Manual 33 Chapter 2 Operation ...
Страница 49: ...moisture IQ User s Manual 37 Chapter 2 Operation ...
Страница 50: ...Chapter 2 Operation 38 moisture IQ User s Manual ...
Страница 98: ...Chapter 6 Setting Up Communications 86 moisture IQ User s Manual no content intended for this page ...
Страница 110: ...Chapter 8 Troubleshooting 98 moisture IQ User s Manual no content intended for this page ...
Страница 138: ...Appendix C Updating the moisture IQ Firmware 126 moisture IQ User s Manual no content intended for this page ...
Страница 180: ... no content intended for this page ...
Страница 181: ......