Chapter 4. Using the Outputs, Alarms and Logger Menus
50
moisture.IQ User’s Manual
4.1.2 Testing the Selected Output
Make sure the output device is connected to the moisture.IQ as described in
complete the following steps to test the output:
1.
Tap the
Channel (1-6)
and
Output
(
A
or
B
) to be tested, and then tap the
Test
button to open a screen
similar to
Figure 44: Output Test Screen
2.
Tap the text box and enter the desired
Percent of Scale
test value (0-120). Then, tap the
Check
button to
send the test value to the output.
3.
After about 5 seconds, this reading should be as shown in
below.
4.
Repeat steps 1-3 for as many different test values as desired. When you have finished testing the
output, tap the
Exit
button to return to the
Output Selection
screen.
Table 7: Expected Multimeter Test Readings
Output Range
Multimeter Reading
0 to 20 mA
20 x Test %/100 mA
4 to 20 mA
4 + 16 x Test %/100 mA
0 to 2V
2 x Test %/100 V
Содержание moisture.IQ
Страница 1: ...Moisture and Oxygen 910 298 Rev E September 2017 bhge com moisture IQ User s Manual ...
Страница 2: ......
Страница 4: ... no content intended for this page ...
Страница 12: ...Preface x moisture IQ User s Manual no content intended for this page ...
Страница 45: ...moisture IQ User s Manual 33 Chapter 2 Operation ...
Страница 49: ...moisture IQ User s Manual 37 Chapter 2 Operation ...
Страница 50: ...Chapter 2 Operation 38 moisture IQ User s Manual ...
Страница 98: ...Chapter 6 Setting Up Communications 86 moisture IQ User s Manual no content intended for this page ...
Страница 110: ...Chapter 8 Troubleshooting 98 moisture IQ User s Manual no content intended for this page ...
Страница 138: ...Appendix C Updating the moisture IQ Firmware 126 moisture IQ User s Manual no content intended for this page ...
Страница 180: ... no content intended for this page ...
Страница 181: ......