![GE moisture.IQ Скачать руководство пользователя страница 120](http://html.mh-extra.com/html/ge/moisture-iq/moisture-iq_user-manual_122539120.webp)
Chapter 9. Specifications
108
moisture.IQ User’s Manual
Sample Requirements
•
Inlet Pressure: 0.2 to 1.0 psig (0.013 to 0.06 barg), standard range
•
Flow Rate: 0.5 to 1.5 SCFH
•
Moisture: No limits (avoid condensation)
•
Oil/Solvent Mist
•
Less than 0.5 mg/ft
3
(standard range)
•
Greater than 0.5 mg/ft
3
(use filter)
•
Solid Particles
•
Less than 2.0 mg/ft
3
(standard range)
•
Greater than 2.0 mg/ft
3
(use filter)