![GE moisture.IQ Скачать руководство пользователя страница 117](http://html.mh-extra.com/html/ge/moisture-iq/moisture-iq_user-manual_122539117.webp)
moisture.IQ User’s Manual
105
Chapter 9. Specifications
9.9 Probe Specifications
Moisture Image Series Probe (MISP or MISP2)
•
Type: Aluminum oxide moisture sensor probe and electronics module
•
Calibration: Each sensor is individually computer calibrated against known moisture concentrations,
traceable to national standards.
•
Process Temperature: -110 to +70°C (-166 to +158°F) dew/frost point temperature
•
Storage Temperature: 70°C (158°F) maximum
•
Operating Pressure: 5
μ
of Hg to 5000 psig
•
Flow Range
•
Gases:
Static to 10,000 cm/sec linear velocity at 1 atm
•
Liquids:
Static to 10 cm/sec linear velocity at 1 g/cc
•
Response Time:
<5 seconds for 63% step change in moisture content in either wet up or dry down cycle
•
Moisture Image Series Probe/Analyzer Separation: 915 m (3000 ft) with provided cable
Note:
Sale of the MISP probe was discontinued by BHGE in 2010 and it was replaced by the MISP2 probe.
Although the MISP probe can be used successfully with the moisture.IQ analyzer, there may be
communication issues when using long cables. Please contact our BHGE technical support team for
assistance.
•
Moisture Image Series Probe/Analyzer Cable: Unshielded, twisted pair, maximum loop resistance of 100
Ω
•
Built-in Temperature Sensor (optional):
•
Type:
Thermistor network
•
Operating Range:
-30 to +70°C (-22 to +158°F)
•
Accuracy: ±0.5°C overall
•
Built-in Pressure Sensor (optional)
•
Type:
Solid state/piezoresistive
•
Available Ranges: 30 to 300 psig, 50 to 500 psig, 100 to 1000 psig, 300 to 3000 psig, 500 to 5000 psig
•
Accuracy:
±1% of span
Содержание moisture.IQ
Страница 1: ...Moisture and Oxygen 910 298 Rev E September 2017 bhge com moisture IQ User s Manual ...
Страница 2: ......
Страница 4: ... no content intended for this page ...
Страница 12: ...Preface x moisture IQ User s Manual no content intended for this page ...
Страница 45: ...moisture IQ User s Manual 33 Chapter 2 Operation ...
Страница 49: ...moisture IQ User s Manual 37 Chapter 2 Operation ...
Страница 50: ...Chapter 2 Operation 38 moisture IQ User s Manual ...
Страница 98: ...Chapter 6 Setting Up Communications 86 moisture IQ User s Manual no content intended for this page ...
Страница 110: ...Chapter 8 Troubleshooting 98 moisture IQ User s Manual no content intended for this page ...
Страница 138: ...Appendix C Updating the moisture IQ Firmware 126 moisture IQ User s Manual no content intended for this page ...
Страница 180: ... no content intended for this page ...
Страница 181: ......