June 2004
1-30
Installation
Standard Delta F
Oxygen Cells
Figure 1-18 below shows the standard oxygen cell and identifies the
sensing and secondary electrodes. Make oxygen cell connections
from the electrodes on the oxygen cell to the OXYGEN terminal
block on the back of the electronics unit. Use Table 1-6 below to
make oxygen cell connections.
IMPORTANT:
To maintain good contact at each terminal block and
to avoid damaging the pins on the connector, pull the
connector straight off (not at an angle), make cable
connections while the connector is away from the
unit, and push the connector straight on (not at an
angle) when the wiring is complete.
Figure 1-18: Standard Delta F Oxygen Cell
Table 1-6: Standard Delta F Oxygen Cell Connections
Connect:
To Delta F
Oxygen Cell:
To Series 1 OXYGEN
Terminal Block
red wire
+ sensing electrode
pin 1
green wire
– sensing electrode
pin 2
white wire
+ secondary electrode
pin 3
black wire
– secondary electrode
pin 4
shield
- - -
pin 5
Secondary
Electrodes
Sensing
Electrodes
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Содержание Moisture Image 1 Series
Страница 8: ...Chapter 1 Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 52: ...Chapter 2 Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 80: ...Chapter 3 Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 96: ...Appendix A Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...