Specifications
3-11
April 2004
M Series Probe
Type
Aluminum oxide moisture sensor probe (patented).
Impedance Range
50 k
Ω
to 2 M
Ω
at 77 Hz (depending on vapor pressure of water).
Calibration
Each sensor is individually computer-calibrated against known
moisture concentrations. Calibrations are traceable to the National
Institute of Standards and Technology (NIST).
Operating Temperature
-110°C to 70°C (-166°F to 158°F)
Storage Temperature
Maximum of 70°C (158°F)
Operating Pressure
(depends on mount)
M1: 5 microns Hg to 75 PSIG
M2: 5 microns Hg to 5000 PSIG
Flow Range
Gases:
From static to 10,000 cm/sec linear velocity at 1 atm
Liquids:
From static to 10 cm/sec linear velocity at density of 1 gm/cc
Response Time:
Less than 5 seconds for 63% of a step change in moisture content in
either wet up or dry down cycle.
Built-in Temperature
Sensor
Type:
Non-linear thermistor
Range:
-30 °C to 70°C (-22°F to 158°F)
Accuracy:
±0.5°C (± 0.33F) overall
Response Time:
Maximum 1 second in well stirred oil.
10 seconds in still air for a 63% step change in increasing or
decreasing temperature.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Содержание Moisture Image 1 Series
Страница 8: ...Chapter 1 Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 52: ...Chapter 2 Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 80: ...Chapter 3 Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 96: ...Appendix A Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...