![GE MG-101 Скачать руководство пользователя страница 19](http://html.mh-extra.com/html/ge/mg-101/mg-101_user-manual_128400019.webp)
MG-101 User’s Manual
11
Chapter 2. Setup and Operation
2.2.2 Using Data Table 3
For calibration points found in
Data Table 3
, refer to
and proceed as follows:
1.
Fully
open the vent valve by turning it counter-clockwise.
CAUTION!
Make the following adjustments slowly, being careful not to overshoot the target value. NEVER
ALLOW THE SYSTEM TO GO TO A VALUE WETTER THAN THE TARGET VALUE. If you overshoot a
value in the wetter direction, you will have to allow the system to equilibrate (dry down) again
before you can proceed.
2.
In
Data Table 3
, look below the body of the table for
Notes 1 and 2
. Adjust both the valves on the
DRY GAS FIRST
DILUTION
and the
DRY GAS SECOND DILUTION
rotameters to the values listed.
IMPORTANT:
Be sure to read the steel ball or glass ball (
S
or
G
), as indicated in the table.
3.
Consult
Data Table 3
for the desired dew/frost point temperature and find the corresponding value in the column
headed
WET GAS
. Adjust the
WET GAS
valve to obtain the value indicated in the table.
4.
By referring to the desired dew/frost point temperature in
Data Table 3
, find the corresponding value in the column
headed
WET GAS DILUTED
. Adjust the vent valve (but not the valve on the
WET GAS DILUTED
rotameter, which must
remain fully open) to obtain the value indicated in
Data Table 3
.
After stabilizing, the system output will be at the selected dew/frost point temperature and it is ready for use. The
stabilization period depends on the dew/frost point temperature you are using. In general, the following times are
required:
•
For dew/frost point temperatures of -60°C (-76°F) and drier, allow 2 hours
•
For dew/frost point temperatures of -60°C (-76°F) to -40°C (-40°F), allow 1 hour
•
For dew/frost point temperatures of -40°C (-40°F) and wetter, allow 30 minutes
After you have completed the calibration or other calibrated moisture operation at this dew/frost point temperature,
either shut down the system (see
), or repeat the steps in this section for the next
calibration point.
Содержание MG-101
Страница 1: ...GE Measurement Control Moisture 910 115 Rev E March 2012 MG 101 User s Manual ...
Страница 2: ......
Страница 4: ...ii no content intended for this page ...
Страница 8: ...Contents vi MG 101 User s Manual no content intended for this page ...
Страница 12: ...Chapter 1 General Information 4 MG 101 User s Manual no content intended for this page ...
Страница 22: ...Chapter 2 Setup and Operation 14 MG 101 User s Manual no content intended for this page ...
Страница 32: ...Warranty 24 MG 101 User s Manual no content intended for this page ...
Страница 33: ......