Recording a Resting ECG
2028451-182B
MAC™ 1600
5-7
Generating a Rhythm Report (Manual Recording)
The
Resting ECG
mode allows you to generate Rhythm Reports, which
are printed reports only. They will not have computer-generated
interpretation or measurements, and they cannot be stored to internal
memory or transmitted. Use the following steps to generate a Rhythm
Report.
1. Prepare the patient as described in Chapter 3.
2. Verify the system is in
Resting ECG
mode.
If the system is not in the
Resting ECG
mode, press the
F1
key at the
Main Menu
to select
Resting ECG
.
3. Enter the patient data as described in Chapter 4.
4. Adjust the
Speed
,
Gain
, and
Low Pass Filter
until the waveforms are
configured as desired.
For more information, refer to
5. If the patient has a pacemaker, press the
F5
key to turn
Pace
Enhance
on.
For more information, refer to
6. Press the
Leads
key to scroll through all 12 leads.
For more information on display formats, refer to
7. Press the
F6
key to select
More
.
8. Press the
F1
key to select the appropriate
Printer Leads
option.
For more information on the Printer Leads option, refer to
9. Press the
Rhythm
key to begin recording the ECG.
10. Press the
Stop
key to stop the ECG recording.
F5
Reanalyze
Allows you to edit the global measurements and t-wave
dispersion. Available only if the
Measurement option is
enabled and the
Reanalysis
option is selected in the
Resting ECG Setup
window
.
For more information, refer to
F6
More
Returns to the setup options.
Refer to
on page 5-4 for details.
F Key
Option
Description
Содержание MAC 1600
Страница 8: ...vi MAC 1600 2028451 182B...
Страница 9: ...2028451 182B MAC 1600 1 1 1 Introduction...
Страница 24: ...1 16 MAC 1600 2028451 182B Introduction...
Страница 25: ...2028451 182B MAC 1600 2 1 2 Equipment Overview...
Страница 38: ...2 14 MAC 1600 2028451 182B Equipment Overview...
Страница 39: ...2028451 182B MAC 1600 3 1 3 Preparing the Patient...
Страница 46: ...3 8 MAC 1600 2028451 182B Preparing the Patient...
Страница 47: ...2028451 182B MAC 1600 4 1 4 Entering Patient Information...
Страница 50: ...4 4 MAC 1600 2028451 182B Entering Patient Information...
Страница 51: ...2028451 182B MAC 1600 5 1 5 Recording a Resting ECG...
Страница 65: ...2028451 182B MAC 1600 6 1 6 Arrhythmia Mode Recording...
Страница 71: ...2028451 182B MAC 1600 7 1 7 Stress Testing...
Страница 82: ...7 12 MAC 1600 2028451 182B Stress Testing...
Страница 83: ...2028451 182B MAC 1600 8 1 8 Managing Internal Storage...
Страница 93: ...2028451 182B MAC 1600 9 1 9 System Configuration...
Страница 128: ...9 36 MAC 1600 2028451 182B System Configuration...
Страница 129: ...2028451 182B MAC 1600 10 1 10 Maintenance...
Страница 143: ...2028451 182B MAC 1600 A 1 A Troubleshooting...
Страница 151: ...2028451 182B MAC 1600 B 1 B Creating Barcodes...
Страница 161: ......