Troubleshooting
2028451-182B
MAC™ 1600
A-3
System Will Not Power Up
If the system will not power up, do the following:
Verify the unit is turned on.
If it is not, turn the unit on. Refer to
page 2-11 for instructions.
Verify the battery is installed and charged.
Refer to
on page A-7 for instructions on verifying
whether the battery is installed and charged.
Refer to
on page 10-12 for instructions on
installing the battery.
Verify the unit is connected to an AC power outlet.
Refer to
“Connecting the AC Power Adapter”
on page 2-8 for
instructions.
Verify the equipment is receiving power from the outlet.
If the unit is receiving power, the Power LED will be lit.
ECG Data Contains Noise
If the acquired ECG data displays unacceptable noise levels, do the
following:
Check the patient’s position.
The patient should remain motionless during the acquisition of a
resting ECG.
Use the
Hookup Advisor
indicator to help determine the cause of the
noise.
For more information, refer to
Verify the electrodes are placed properly.
Refer to
on page 3-3 for information on
proper electrode placement.
Verify the electrodes have been applied correctly.
Perspiration, excessive hair, lotions, and dead skin cells must be
removed from the electrode site.
Refer to
on page 3-2 more information.
Check for defective or expired electrodes.
Replace the electrodes if there are any questions about their
effectiveness.
Check for defective, broken, or disconnected leadwires.
Replace the leadwires if there are any questions about their
effectiveness. Refer to
Consider using filters,
ADS
, and
FRF
to help eliminate or reduce
ECG noise.
For more information, refer to
on page 6-3, or
Содержание MAC 1600
Страница 8: ...vi MAC 1600 2028451 182B...
Страница 9: ...2028451 182B MAC 1600 1 1 1 Introduction...
Страница 24: ...1 16 MAC 1600 2028451 182B Introduction...
Страница 25: ...2028451 182B MAC 1600 2 1 2 Equipment Overview...
Страница 38: ...2 14 MAC 1600 2028451 182B Equipment Overview...
Страница 39: ...2028451 182B MAC 1600 3 1 3 Preparing the Patient...
Страница 46: ...3 8 MAC 1600 2028451 182B Preparing the Patient...
Страница 47: ...2028451 182B MAC 1600 4 1 4 Entering Patient Information...
Страница 50: ...4 4 MAC 1600 2028451 182B Entering Patient Information...
Страница 51: ...2028451 182B MAC 1600 5 1 5 Recording a Resting ECG...
Страница 65: ...2028451 182B MAC 1600 6 1 6 Arrhythmia Mode Recording...
Страница 71: ...2028451 182B MAC 1600 7 1 7 Stress Testing...
Страница 82: ...7 12 MAC 1600 2028451 182B Stress Testing...
Страница 83: ...2028451 182B MAC 1600 8 1 8 Managing Internal Storage...
Страница 93: ...2028451 182B MAC 1600 9 1 9 System Configuration...
Страница 128: ...9 36 MAC 1600 2028451 182B System Configuration...
Страница 129: ...2028451 182B MAC 1600 10 1 10 Maintenance...
Страница 143: ...2028451 182B MAC 1600 A 1 A Troubleshooting...
Страница 151: ...2028451 182B MAC 1600 B 1 B Creating Barcodes...
Страница 161: ......