background image

GE 

LOGIQ e/LOGIQ e Vet/LOGIQ i/
Vivid e

Basic Service Manual 

Part Number: 5370626-100

Revision: 18

Содержание LOGIQ e R4 Series

Страница 1: ...GE LOGIQ e LOGIQ e Vet LOGIQ i Vivid e Basic Service Manual Part Number 5370626 100 Revision 18...

Страница 2: ...This manual covers the software versions for the products as below LOGIQ e R4 x x R5 x x R6 x x R7 x x LOGIQ e Vet R5 x x R6 x x R7 x x VIVID e R4 x x R5 x x R6 x x LOGIQ i R4 x x R5 x x...

Страница 3: ...GUE QUE L ANGLAIS C EST AU CLIENT QU IL INCOMBE DE LE FAIRE TRADUIRE NE PAS TENTER D INTERVENTION SUR LES QUIPEMENTS TANT QUE LE MANUEL SERVICE N A PAS T CONSULT ET COMPRIS LE NON RESPECT DE CET AVERT...

Страница 4: ...DOR OU PACIENTE DEVIDO A CHOQUES EL TRICOS MEC NICOS OU OUTROS ESTE MANUAL DE ASSIST NCIA EST DISPON VEL APENAS EM INGL S SE QUALQUER OUTRO SERVI O DE ASSIST NCIA T CNICA QUE N O A GEHC SOLICITAR ESTE...

Страница 5: ...NINKIELIST MATERIAALIA TARVITTAVAN K NN KSEN HANKKIMINEN ON ASIAKKAAN VASTUULLA L YRIT KORJATA LAITTEISTOA ENNEN KUIN OLET VARMASTI LUKENUT JA YMM RT NYT T M N HUOLTO OHJEEN MIK LI T T VAROITUSTA EI N...

Страница 6: ...U EB OBSLU N HO PERSON LU NEBO PACIENT VLIVEM ELEKTRICK HOP PROUDU RESPEKTIVE VLIVEM K RIZIKU MECHANICK HO PO KOZEN NEBO JIN MU RIZIKU DENNE SERVICEMANUAL FINDES KUN P ENGELSK HVIS EN KUNDES TEKNIKER...

Страница 7: ...TAS GALI B TI SU EISTAS D L ELEKTROS SM GIO MECHANINI AR KIT PAVOJ DENNE SERVICEH NDBOKEN FINNES BARE P ENGELSK HVIS KUNDENS SERVICELEVERAND R TRENGER ET ANNET SPR K ER DET KUNDENS ANSVAR S RGE FOR OV...

Страница 8: ...I AVERTISMENT AR PUTEA DUCE LA R NIREA DEPANATORULUI OPERATORULUI SAU PACIENTULUI N URMA PERICOLELOR DE ELECTROCUTARE MECANICE SAU DE ALT NATUR OVAJ PRIRU NIK ZA SERVISIRANJE DOSTUPAN JE SAMO NA ENGLE...

Страница 9: ...LI PACIJENTA PROUZRO ENE STRUJNIM UDAROM MEHANI KIM I DRUGIM OPASNOSTIMA T TO SERVISN PR RU KA JE K DISPOZ CII LEN V ANGLI TINE AK Z KAZN KOV POSKYTOVATE SLU IEB VY ADUJE IN JAZYK AKO ANGLI TINU POSKY...

Страница 10: ...SYEN KILAVUZUN NG L ZCE DI INDAK B R D LDE OLMASINI STERSE KILAVUZU TERC ME ETT RMEK M TER N N SORUMLULU UNDADIR SERV S KILAVUZUNU OKUYUP ANLAMADAN EK PMANLARA M DAHALE ETMEY N Z BU UYARININ G Z ARDI...

Страница 11: ...GE Direction 5370626 100 Revision 18 LOGIQ e LOGIQ e Vet LOGIQ i Vivid e Basic Service Manual ix ZH CN KO...

Страница 12: ...will comply with the requirements of the applicable electrical codes The purchaser of GE equipment shall only utilize qualified personnel i e GE s field engineers personnel of third party service com...

Страница 13: ...duplicated in any form in whole or in part without prior written permission of GE Healthcare GE Healthcare may revise this publication from time to time without written notice TRADEMARKS All products...

Страница 14: ...2012 Add instruction for Loading the System Software with USB Memory Stick 10 10 Dec 2012 Update Chapter 9 about spare parts 11 17 Dec 2012 Update package label and add new spare parts 12 23 Jan 2013...

Страница 15: ...1 6 Important Conventions 1 7 Conventions Used in Book 1 7 Standard Hazard Icons 1 8 Product Icons 1 9 Safety Considerations 1 13 Introduction 1 13 Human Safety 1 13 Mechanical Safety 1 13 Electrical...

Страница 16: ...pose of this chapter 2 2 1 Chapter Contents 2 1 General Console Requirements 2 2 Console Environmental Requirements 2 2 Electrical Requirements 2 2 EMI Limitations 2 4 Scan Probe Environmental Require...

Страница 17: ...Order 3 8 Physical Inspection 3 8 EMI Protection 3 8 Completing the Installation 3 9 Power On Boot Up 3 9 Power Off Shutdown 3 10 Transducer Connection 3 11 System Configuration 3 12 System Specificat...

Страница 18: ...4 3 Archiving and Loading Presets 4 5 Adjusting the Display Monitor 4 6 Lockout Tagout LOTO requirements 4 6 System Features 4 7 B Mode Checks 4 10 M Mode Controls 4 13 Color Flow Mode Checks 4 16 Do...

Страница 19: ...2 General Information 5 3 External I O 5 4 Peripherals 5 4 Wiring 5 5 Power Diagrams 5 6 Overview 5 6 AC Power 5 6 Battery charging 5 7 Air Flow Distribution 5 8 Fans 5 9 Common Service Platform 5 10...

Страница 20: ...6 100 Revision 18 LOGIQ e LOGIQ e Vet LOGIQ i Vivid e Basic Service Manual Table of Contents 6 CHAPTER 6 Service Adjustments Overview 6 1 Purpose of this chapter 6 6 1 Monitor Adjustments 6 2 Adjustme...

Страница 21: ...s 7 6 Check and Record the P1 Key Function 7 6 Setting the P1 Key to Screen Capture 7 6 Capturing a Screen 7 6 Reset the P1 Key to Customer s Functionality 7 8 Global Service User Interface GSUI 7 9 E...

Страница 22: ...e LOGIQ e Vet LOGIQ i Vivid e 8 1 Warning and Caution 8 1 Handle Assy FRU No 313 8 2 Loading the System Software 8 4 Loading Base Image Software with DVD 8 5 Loading the System Software with USB Memo...

Страница 23: ...1 List of Abbreviations 9 1 Renewal Parts Lists 9 2 Equipment Models Covered in this Chapter 9 2 Operator Console Assy 9 3 LCD Assy 9 4 Keyboard Assy 9 7 Bottom Assy 9 11 Options of spare parts for L...

Страница 24: ...be performed 10 3 Tools Required 10 4 Special Tools Supplies and Equipment 10 4 System Maintenance 10 5 Preliminary Checks 10 5 Functional Checks See Also Chapter 4 10 6 Input Power 10 7 Cleaning 10...

Страница 25: ...ctional checks that are recommended as part of the installation or as required during servicing and periodic maintenance 5 Chapter 5 Components and Functions Theory Contains block diagrams and functio...

Страница 26: ...OGIQ e LOGIQ e Vet LOGIQ i Vivid e Basic Service Manual 1 2 1 1 4 Typical Users of the Basic Service Manual Service Personnel installation maintenance etc Hospital s Service Personnel Contractors Some...

Страница 27: ...Q e Vet R5 0 x Console for Europe 5198793 LOGIQ e LOGIQ e Vet R5 0 x Console for China 5198161 LOGIQ e LOGIQ e Vet R5 0 x Console for Japanese 5199693 LOGIQ e LOGIQ e Vet R5 0 x Console for Australia...

Страница 28: ...x Console for India 5432754 LOGIQ e R7 x x Console for Korea 5439035 LOGIQ e R7 x x Console for CKD USA 5447891 LOGIQ e R7 x x Console for China Table 1 7 LOGIQ i R4 x x Model Designations Part Numbe...

Страница 29: ...13329 Vivid e R5 x x Console for Japan 5212851 Vivid e R5 x x Console for Australia 5199644 Vivid e R5 x x Console for India South Africa 5183318 Vivid e R5 x x Console with SKD ECG for Europe 5435024...

Страница 30: ...Vivid e R6 x x Console for EU 5435035 Vivid e R6 x x Console for Argentina 5435036 Vivid e R6 x x Console for India 5435037 Vivid e R6 x x Console for Australia 5435038 Vivid e R6 x x Console for Jap...

Страница 31: ...r potential hazards are labeled in one of following ways NOTE Notes provide important information about an item or a procedure Information contained in a NOTE can often save you time or effort DANGER...

Страница 32: ...ter In addition to text several different graphical icons symbols may be used to make you aware of specific types of hazards that could cause harm Other hazard icons make you aware of specific procedu...

Страница 33: ...indicate the degree of safety or protection Bottom panel of the adapter IP Code IPX1 or IPX8 IPX1 FSU 2001 IPX8 MKF 2 MED GP26 FSU 1000 Indicates the degree of protection provided by the enclosure pe...

Страница 34: ...e and or logo of the testing laboratory product category safety standard is assessed and a control number Bottom panel of the console Date of manufacture The date could be a year year and month or yea...

Страница 35: ...to automatically close Rating Plate Indicates the product contains hazardous materials in excess of the limits established by Chinese standard SJ T11363 2006 Requirements for Concentration Limits for...

Страница 36: ...battery may be safely removed from the device please consult the service manual or equipment instructions Information on the potential effects on the environment and human health of the substances us...

Страница 37: ...unpaved roads excessive speeds and erratic stops or starts 1 3 4 Electrical Safety To minimize shock hazard the equipment chassis must be connected to an electrical ground The system is equipped with...

Страница 38: ...the battery causing equipment damage observe the following precautions Do not immerse the battery in water or allow it to get wet Do not put the battery into a microwave oven or pressurized container...

Страница 39: ...ARE PRESENT IN THIS EQUIPMENT USE EXTREME CAUTION WHEN HANDLING TESTING AND ADJUSTING WARNING WARNING EXPLOSION WARNING DO NOT OPERATE THE EQUIPMENT IN AN EXPLOSIVE ATMOSPHERE OPERATION OF ANY ELECTRI...

Страница 40: ...or intended for use in patient care are regulated medical waste for transportation purposes and must be transported as a hazardous material NOTE The USER SERVICE staff should dispose all the waste pro...

Страница 41: ...cables In addition to electromagnetic energy EMC also includes possible effects from electrical fields magnetic fields electrostatic discharge and disturbances in the electrical power supply 1 4 2 CE...

Страница 42: ...Ultrasound Service Engineering 9900 Innovation Drive Wauwatosa WI 53226 Service On site Service Parts Applications support 1 800 437 1171 1 800 558 2040 1 800 682 5327 or 1 262 524 5698 14470 Bergen B...

Страница 43: ...8 TEL 86 510 85225888 FAX 86 510 85226688 Table 1 19 Factory Sites Manufacturer Phone Number GE Medical Systems China Co Ltd No 19 Changjiang Road Wuxi National Hi Tech Dev Zone Jiangsu P R China 2140...

Страница 44: ...GE Direction 5370626 100 Revision 18 LOGIQ e LOGIQ e Vet LOGIQ i Vivid e Basic Service Manual 1 20 Section 1 5 Customer Assistance This page was intentionally left blank...

Страница 45: ...ter 2 This chapter provides the information required to plan and prepare for the installation of a LOGIQ e LOGIQ e Vet LOGIQ i Vivid e Included are descriptions of the facility and electrical needs to...

Страница 46: ...d line shall consist of one phase a neutral not shared with any other circuit and a full size ground wire from the distribution panel to the Ultrasound outlet Sites with a mains power system without a...

Страница 47: ...est equipment needed to support this unit must also be present within 1 m 3 2 ft of the unit Electrical installation must meet all current local state and national electrical codes 2 2 2 5 Unit Power...

Страница 48: ...oor grounding is the most likely reason a unit will have noisy images Check grounding of the power cord and power outlet Replace all screws RF gaskets covers cores After you finish repairing or updati...

Страница 49: ...0 C Storage 10 to 50 C NOTE The recommended storage temperature range of 6Tc RS is between 0 C and 45 C NOTE Temperature in degrees C Conversion to Degrees F Degrees C 9 5 32 NOTICE SYSTEMS AND ELECTR...

Страница 50: ...ed electrical contractors Other connections between pieces of electrical equipment products involved and the accompanying electrical installations are highly sophisticated and special engineering comp...

Страница 51: ...trasound unit is only a conduit Dedicated single branch power outlet of adequate amperage meeting all local and national codes which is located less than 2 5 m 8 ft from the unit s proposed location D...

Страница 52: ...OVE R HE AD LI G HT S DIMME R DE DIC AT E D POWE R R E C EP TA C LE E XT E R NAL PE R IP HE R ALS SIN K C O UN T E R TO P FIL M VIE WE R FIL M SUPPLIES SIN K LINEN SUPPLY PR OB ES SUPPLIES 24 IN 61 C...

Страница 53: ...tor and peripherals enabling viewing to be done while scanning continues With DICOM images can be archived stored and retrieved faster easier and at a lower cost 2 3 4 4 DICOM Option Pre installation...

Страница 54: ...ents cont d Figure 2 2 Worksheet for DICOM Network Information LOGIQ e Host Name AE Title Local Port IP Address Net Mask ROUTING INFORMATION ROUTER2 ROUTER3 ROUTER1 Destination IP Addresses GATEWAY IP...

Страница 55: ...tion and how to check and test the unit probes and external peripherals for electrical safety are included in this procedure Also included in this section are guidelines for transporting the unit to a...

Страница 56: ...functional checkout will take approximately one hour LOGIQ e LOGIQ e Vet LOGIQ i Vivid e consoles with optional equipment may take slightly longer Table 3 2 Average Installation Time Description Aver...

Страница 57: ...5 20 15 10 5 0 5 10 15 20 25 30 35 40 F 140 131 122 113 104 95 86 77 68 59 50 41 32 23 14 5 4 13 22 31 40 hrs 8 6 4 2 0 0 0 0 0 0 0 2 4 6 8 10 12 14 16 18 20 DANGER WHEN USING ANY TEST INSTRUMENT THAT...

Страница 58: ...ng and Unpacking the Equipment When a new system arrives check that any components are not damaged and are not in short supply If shipping damage or shortage occurs contact the address shown in Chapte...

Страница 59: ...cont d 3 Take out the Paper pad 4 Take out console together with 2 interleavers from console package 5 Take out the interleavers beside Accessories Package 6 Take out Accessories Package Figure 3 2 U...

Страница 60: ...Vivid e Basic Service Manual 3 6 Section 3 2 Receiving and Unpacking the Equipment Section 3 2 Receiving and Unpacking the Equipment cont d 7 Remove 2 interleavers 8 Remove plastic bag Figure 3 3 Remo...

Страница 61: ...GIQ e Vet LOGIQ i Vivid e Before moving store all loose parts in original accessory box or in back pack Return probes to original box Section 3 3 Packing the Equipment Please pack LOGIQ e LOGIQ e Vet...

Страница 62: ...ngs for the LOGIQ e LOGIQ e Vet LOGIQ i Vivid e Scanner is found on a label located on the AC adapter 220 240VAC China 100 120VAC USA Japan 220 240VAC Europe Latin America 3 4 3 EMI Protection This Un...

Страница 63: ...stem may not be able to boot if power is recycled too quickly 3 5 1 1 Scanner Power On Lower the handle Plug the AC adapter output connector into the system DC input port located on the system s rear...

Страница 64: ...turning off a system wait at least ten seconds before turning it on again The system may not be able to boot if power is recycled too quickly 3 5 2 1 Back end Processor Power Down To power down the sy...

Страница 65: ...mple Relocating the scanner 3 5 3 Transducer Connection 1 Carefully open the system LCD display plug the probe connector into the probe port then lock the probe latch upward NOTE Please ensure that th...

Страница 66: ...s of the LOGIQ e LOGIQ e Vet LOGIQ i Vivid e console with old LCD are summarized in Figure 3 9 on page 3 12 Table 3 4 Physical Dimensions of LOGIQ e LOGIQ e Vet LOGIQ i Vivid e with old LCD Height Wid...

Страница 67: ...Vivid e console with new LCD are summarized in Figure 3 10 on page 3 13 Table 3 5 Physical Dimensions of LOGIQ e LOGIQ e Vet LOGIQ i Vivid e with new LCD Height Width Depth Unit Console Only Console w...

Страница 68: ...and new LCD Identify LCD type from the appearance of Ultrasound System 3 6 2 Electrical Specifications Figure 3 11 Difference between Old LCD and New LCD Table 3 6 Electrical Specifications for LOGIQ...

Страница 69: ...120 USB LITEON LITEON Model eHAU 324 USB PLEXTOR PX L890UE USB LITEON LITEON eSAU108 DVDRW USB LITEON LITEON eUAU108 DVDRW USB LITEON LITEON eBAU108 DVDRW USB Wireless Lan Adapter NetGear WNA3100 USB...

Страница 70: ...800U USB Not supported by LOGIQ e R4 x x Vivid e R4 x x LOGIQ i R4 x x Video Transfer Pinnacle Pinnacle Video Transfer 8230 10022 11 USB Not supported by LOGIQ e R4 x x 5 0 x LOGIQ e Vet R5 0 x Vivid...

Страница 71: ...Connector Located on the rear panel are video input and output connectors power connector and ethernet port 1 Port for DC input AC Adapter 2 Docking Connector 3 SVGA Output 4 Ethernet port WARNING WAR...

Страница 72: ...d on the left side panel are two USB Ports for control connections for printer service tools and earphone port 1 Two interchangeable USB Port Digital Printers DVD RW and or FootSwitch etc 2 Earphone P...

Страница 73: ...USB Table 3 10 Pin assignment of USB1 Table 3 11 Pin assignment of USB2 3 Pin assignment of RS232C for external VGA Connector D SUB 15Pin Female Table 3 12 Pin Assignments of RS232C for External VGA...

Страница 74: ...using USB Port1 or USB Port 2 Figure 3 14 on page 3 20 B Connect UP D23MD color printer to the system UP D23MD Color Printer can be properly connected using USB Port1 or USB Port 2 Figure 3 15 on page...

Страница 75: ...ve worked successfully before Overall system configuration and other factors may also affect the success of creating a DVD CD The following media types have been tested LITEON highly recommends that y...

Страница 76: ...tch can be properly connected using USB Port1 or USB Port2 E Connect Wireless LAN Adapter to the system Wireless LAN Adapter can be properly connected using USB port F Connect the CRT to the system CR...

Страница 77: ...apter 3 System Setup 3 23 3 6 5 4 Connect peripherals cont d G Connect the USB Memory to the system The USB Memory can be properly connected using USB port 1 or 2 H Connect the ECG to the system ECG c...

Страница 78: ...ation 3 6 5 4 Connect peripherals cont d I Connect the USB HDD to the system The USB Harddisk can be properly connected using USB port 1 and 2 J Connect the BarCode Reader to the system the BarCode Re...

Страница 79: ...disk U3 USB Flash Drive is Predefined NOTE The Sandisk U3 USB Flash Drive is only available on LOGIQ e R6 x x R7 x x and LOGIQ e Vet R6 x x R7 x x Connect the security enabled Sandisk U3 USB Flash Dri...

Страница 80: ...sh Drive will be locked if reaching the maximum number of failed password attempts NOTE Please refer to the operation manual of each peripheral for information needed by the user to operate the system...

Страница 81: ...ay box and select OK NOTE Please refer to the operation manual of each peripheral for information needed by the user to operate the system safely NOTE For LOGIQ e R5 2 x R6 x x R7 x x LOGIQ e Vet R6 x...

Страница 82: ...x the system supports 4C RS E8C RS 8C RS i12L RS 8L RS 3S RS 6S RS 12L RS i t739 RS P2D 9L RS Probes For Vivid e R4 x x the system supports 4C RS i12L RS 8L RS 3S RS Probes For Vivid e R5 0 x the sys...

Страница 83: ...TIVE SMALL PARTS VASCULAR PEDIATRICS LINEAR H40402LW 2377942 8L RS VALOX SMALL PARTS PERIPHERAL VASCULAR LINEAR H40402LT 2376127 3S RS VALOX CARDIOLOGY TRANSVAGINAL ABDOMEN SECTOR H4000PD 2355686 12L...

Страница 84: ...r 16 Customizing Your System Table 3 15 List of Probes for LOGIQ e Vet Probe Name Material of Headshell Area of Using TYPE Catalog Number Part Number 4C RS Vet NORYL GENERAL PURPOSE CONVEX H40442LY 51...

Страница 85: ...te System Information Services Destination Devices TCP IP Settings Device Type Manufacturer Name IP Address Port AE Title 1 2 3 4 5 6 7 8 9 10 11 12 Name AE Title Dept LOGIQ SN Type Floor Room REV Com...

Страница 86: ...0 Revision 16 LOGIQ e LOGIQ e Vet LOGIQ i Vivid e Basic Service Manual 3 32 Section 3 9 Loading Base Image Software Section 3 9 Loading Base Image Software Please refer to Section 8 3 Loading the Syst...

Страница 87: ...re Version check out 3 10 1 Functional Check out 1 Power on LOGIQ e LOGIQ e Vet LOGIQ i Vivid e scanner and wait until system booting to main screen 2 Press Utility Config key on control panel 3 Choos...

Страница 88: ...levant safety and user information is available during the operation and service of the complete system 3 11 1 Product Locator Installation NOTE The Product Locator Installation Card shown may not be...

Страница 89: ...IQ e Vet LOGIQ i Vivid e console diagnostics by using the built in service software and power supply adjustments Section 4 2 Required Equipment To perform these tests you ll need any of the sector lin...

Страница 90: ...utput connector into the system DC input port located on the system s rear panel with the arrow side upward Plug the AC adapter power cord into a grounded protective earth outlet CAUTION SYSTEM REQUIR...

Страница 91: ...pressed once the Back end Processor starts and the software code is distributed to initiate the scanner No status messages are displayed during this process 4 3 2 Power Off Shutdown NOTE After turnin...

Страница 92: ...tenance Reboot Full Maintenance Reboot will fully restart the whole system for performance improvement NOTE It s recommended to do Full Maintenance Reboot at least daily or when system starts to slow...

Страница 93: ...stem 4 3 3 1 Archiving Presets to an DVD R Disk or USB memory device 1 Insert an empty blank DVD R disk into the DVD RW 2 Access to the Utility Config Menu and select System The Backup sheet will be s...

Страница 94: ...Tagout requirements by ensuring you are in total control of the AC power plug at all times during the service process To apply Lockout Tagout 1 Plan and prepare for shutdown 2 Shutdown the equipment...

Страница 95: ...Chapter 4 Functional Checks 4 7 4 3 6 System Features 4 3 6 1 Control Panel 1 TGC 2 New Patient 3 End Exam 4 Mode Gain Auto Keys 5 Preset Key 6 Imaging Measurement Keys 7 Depth Zoom Ellipse 8 Image K...

Страница 96: ...i Vivid e SoftMenu Key Tour In general there are two types of softMenu keys Paddle Switch and adjustable knobs 1 The Paddle Switch is used to access and adjust the Sub SoftMenu 2 Press the adjustable...

Страница 97: ...3 Acoustic Output Readout 15 TGC not shown on the image 4 GE Symbol Probe Orientation Marker Coincides with a probe orientation marking on the probe 16 Body Pattern 5 Image Preview 17 Depth Scale 6 G...

Страница 98: ...3 6 6 Available Probes on page 3 28 in Chapter 3 Installation to the System probe connector 2 Turn ON the scanner if it isn t turned on already NOTE The keyboard layout varies from each Product the f...

Страница 99: ...the beam for specific area Press the control to toggle between Focus Position and Focus Number Press Up Down Button to move or adjust the focal numbers 5 Activate Auto Optimize Optimize the image bas...

Страница 100: ...r shift up to a higher frequency 6 Adjust Frame Average Temporal filter that averages frames together This has the effect of presenting a smoother softer image For LOGIQ e LOGIQ e Vet and LOGIQ i Adju...

Страница 101: ...ntrols 4 3 8 1 Preparations 1 Connect one of the probes listed in 3 6 6 Available Probes on page 3 28 in Chapter 3 Installation to the System probe connector 2 Turn ON the scanner if it isn t turned o...

Страница 102: ...anel Controls Step Task Expected Result s Comments 1 Press M Mode key M Mode Starts 2 Adjust Gain Controls the amount of echo information displayed in an image Turn B Mode dial to the left right to in...

Страница 103: ...ries by enhancing the gray scale differences corresponding to the edges of structures Adjustments to M Mode s edge enhancement affects the M Mode only For Vivid e 4 Activate Gray Map Determines how th...

Страница 104: ...de Checks 4 3 9 1 Preparations 1 Connect one of the probes listed in 3 6 6 Available Probes on page 3 28 in Chapter 3 Installation to the System probe connector 2 Turn ON the scanner if it isn t turne...

Страница 105: ...3 9 2 Color Flow Mode OP Panel Controls Table 4 7 Color Flow Mode OP Panel Controls Step Task Expected Result s comments 1 Press CFM Mode key CFM Mode Starts 2 Adjust Gain Amplifies the overall stren...

Страница 106: ...t perspective Press Invert to reverse the color map 7 Adjust LIne Density Trades frame rate for sensitivity and spatial resolution If the frame rate is too slow reduce the size of the region of intere...

Страница 107: ...to select specific transparency map 15 Focus Position Increases the number of transmit focal zones or moves the focal zone s so that you can tighten up the beam for a specific area A graphic caret co...

Страница 108: ...10 1 Preparations 1 Connect one of the probes listed in 3 6 6 Available Probes on page 3 28 in Chapter 3 Installation to the System probe connector 2 Turn ON the scanner if it isn t turned on already...

Страница 109: ...in Amplifies the overall strength of the echoes processed in the Color Flow window Turn the Gain dial PW Mode key to the left right to increase decrease Gain 3 Display M D Mode Cursor Displays the M D...

Страница 110: ...trace direction 7 Invert Vertically inverts the spectral trace without affecting the baseline position Press invert to invert the spectral trace The Plus and Minus signs on the velocity scale reverse...

Страница 111: ...ngle gate PRF capability the system automatically switches to high PRF mode Multiple gates appear and HPRF is indicated on the display For Vivid e 18 Trace Sensitivity Adjusts the sensitivity to get m...

Страница 112: ...IQ i press F6 on the keyboard to enter CW mode F6 is the predefined function key for CW which can be configured in Utility Config Admin Function Key preset menu For LOGIQ e R7 x x and LOGIQ e Vet R7 x...

Страница 113: ...oint press SET The system fixes the first caliper and displays a second active caliper 4 To position the second caliper move the TRACKBALL 5 Adjust the ELLIPSE an ellipse with an initial circle shape...

Страница 114: ...e in its storage box 4 3 14 Using Cine 4 3 14 1 Activating CINE Press Freeze then roll the Trackball to activate CINE To start CINE Loop playback press Run Stop To stop CINE Loop playback press Run St...

Страница 115: ...move through CINE memory one frame at a time 4 3 15 Image Management QG For Image Management functionality refer to the LOGIQ e LOGIQ e Vet LOGIQ i Vivid e Quick Guide It talks about several topics C...

Страница 116: ...he backup media 3 Select the media type from the drop down menu 4 Enter the label for the media as shown in Figure 4 15 It is best to use all capital letters with no spaces or punctuation marks Press...

Страница 117: ...e drive 2 On the Keyboard press UTILITY CONFIG 3 On the LCD display press SYSTEM 4 On the LCD display select BACKUP RESTORE NOTE If you are not logged in as GE Service or with administrator privileges...

Страница 118: ...window is displayed Log on with administrator privileges 5 In the Restore list select Patient Archive Report Archive and User Defined Configuration 6 In the Media field select the Backup Restore DVD R...

Страница 119: ...abel media by date NOTE Images will be moved from the hard disk drive by date Therefore the best way to label media is by date When images are moved to the archive media they will be deleted from the...

Страница 120: ...4 Archiving Images cont d 7 Press PATIENT and set the Dataflow to store images directly to DVD ROM 8 From the image screen press EZBACKUP MOVE The Move Images pop up appears 9 Press Next on pop up me...

Страница 121: ...Press Finish after Backup Move complete All databases presets and images should now be saved to removable media 4 3 16 5 Emergency Repair Disk 1 Connect the USB HDD to system 2 Select Backup Restore t...

Страница 122: ...ral Procedure 4 3 16 5 Emergency Repair Disk cont d 4 Return to scan mode Press Ctrl B the Emergency Disk Making window pop up appears 5 Press Next on pop up message 6 A storage information message ap...

Страница 123: ...35 4 3 16 5 Emergency Repair Disk cont d 7 Press Finish after Emergency Disk Making complete All databases presets and images should now be saved to Emergency Disk 8 Press Ctrl R Emergency Disk Recove...

Страница 124: ...Manual 4 36 Section 4 3 General Procedure 4 3 16 5 Emergency Repair Disk cont d 10 A recovery information message appears Press Next to continue 11 Press Finish after Emergency Disk Recovery complete...

Страница 125: ...ment of Printer Keys For LOGIQ e LOGIQ e Vet LOGIQ i the default function for Print1 3 Keys is P1 store image P2 print P3 USB Quick Save Print1 3 Keys can also be assigned as desired by the customer F...

Страница 126: ...GE Direction 5370626 100 Revision 18 LOGIQ e LOGIQ e Vet LOGIQ i Vivid e Basic Service Manual 4 38 Section 4 5 Peripheral Checks This page was intentionally left blank...

Страница 127: ...5 1 Overview This chapter explains LOGIQ e LOGIQ e Vet LOGIQ i Vivid e s system concepts component arrangement and subsystem function It also describes the Power Distribution System PDS and probes Tab...

Страница 128: ...t DC DC HV Ctrl SMBUS I F DC DC high voltage control smart bus interface TUSC FPGA LOGIQ e LOGIQ e Vet LOGIQ i Vivid e Ultrasound controlling and Image data transfer FPGA Figure 5 1 LOGIQ e LOGIQ e Ve...

Страница 129: ...ging 2D Color Flow M Mode Black and White imaging Doppler A number of combinations of the above LOGIQ e LOGIQ e Vet LOGIQ i Vivid e is a digital beam forming system that can handle up to 128 elements...

Страница 130: ...External I O External I O The external I O is the interface between the scanner and all external items Examples wireless network USB interface medical grade printer and external medical grade SVGA di...

Страница 131: ...e Vet LOGIQ i Vivid e Basic Service Manual Chapter 5 Components and Functions Theory 5 5 5 2 5 Wiring TX64 Transmit board RX64 Front end processing board TMST Master board PWR SW Power Switch FPC Flex...

Страница 132: ...no main power switch located on this power pack 5 3 2 AC Power The mains cord has plugs in one side end A male plug connects to the mains outlet on site The mains voltage is routed to the AC power pa...

Страница 133: ...the battery and the output of AC Pack If the output of AC Pack is available the power input of Charge Board Unit should be from the AC Pack and the battery will be charged if it s not full This block...

Страница 134: ...Power Diagrams 5 3 4 Air Flow Distribution The two air flow passes allow the scanner to be cooled down as shown in the figure above Path A Bottom front CPU Assy Bottom left for TMST CPU Assy cooling P...

Страница 135: ...ice Manual Chapter 5 Components and Functions Theory 5 9 5 3 5 Fans The scanner contains the four fans at the following positions for producing an air flow One fan assy On the Left side for air flow p...

Страница 136: ...raining and service costs 5 4 2 Global Service User Interface GSUI 5 4 2 1 Internationalization The user interface provided by the service platform is designed for GE personnel and as such is in Engli...

Страница 137: ...le around the immovable object refer to Figure 5 9 on page 5 11 2 Make sure and rotate the key to the right unlocked position 3 Insert the lock into the Kensington security slot in the system side cov...

Страница 138: ...e Vet LOGIQ i Vivid e Basic Service Manual 5 12 Section 5 4 Common Service Platform 5 4 3 Service Home Page The navigation bar at the top of the screen allows the user to select from several tools an...

Страница 139: ...n sub menus and an Exit selection The Service Interface allows scanner logs to be viewed by all service users The log entries are color coded to identify the error level severity at a glance The Servi...

Страница 140: ...lculationMade CalculationTime HelpUsage ProductName SystemSWRev NoOfSWModules SystemPowerOnTime SystemIdleTime NoOfBoards ProbeChangedDate ProbeChangedTime ProbeChangedStatus NoOfProbeSlots NoOfProbes...

Страница 141: ...Log contents with the package on the x axis and incident count on the y axis Plot Page Allows for the color coded plot of all Page contents with the package on the x axis and incident count on the y...

Страница 142: ...ision 18 LOGIQ e LOGIQ e Vet LOGIQ i Vivid e Basic Service Manual 5 16 Section 5 4 Common Service Platform 5 4 4 4 Exit The sub menu Exit Log Viewer returns the user to the Service Desktop home page F...

Страница 143: ...s are executable by both local and remote users The Service Platform provides top level diagnostic selection based on the user s level and login access permissions Remote access will require disruptiv...

Страница 144: ...orm 5 4 6 Image Quality The Image Quality page is intended to contain tools for troubleshooting image quality issues 5 4 7 Calibration The Calibration page is intended to contain the tools used to cal...

Страница 145: ...ion page is intended to be used to setup various configuration files on the system The Service Platform is the access and authorization control for remote access to the configuration subsystem The ena...

Страница 146: ...Replacement The Replacement page is intended to contain the tools used to track replacement parts used in the system 5 4 11 PM The PM page is intended to contain the tools used in periodic maintenanc...

Страница 147: ...6 1 Chapter 6 Service Adjustments Section 6 1 Overview 6 1 1 Purpose of this chapter 6 This section describes how to test and adjust the scanner These tests are optional You may use them to check the...

Страница 148: ...6 2 Section 6 2 Monitor Adjustments Section 6 2 Monitor Adjustments 6 2 1 Adjustments Procedures To adjust the brightness and volume 1 Press Ctrl brightness key together up down located on the keyboa...

Страница 149: ...maintain image quality and system operation Very basic host system and board level diagnostics are run whenever power is applied Some Service Tools may be run at the application level However most sof...

Страница 150: ...k to the manufacturer for analysis There are different options to acquire this data that would give different results 7 2 2 Collect Vital System Information The following information is necessary in o...

Страница 151: ...Crash Log Power Supply Temperature NOTE Power Supply and Temperature logs are not currently being updated by the LOGIQ e LOGIQ e Vet LOGIQ i Vivid e This Alt D function is available at all times When...

Страница 152: ...e However the default is for the Image Area only or the customer may have customized the default Key function Here take LOGIQ e for example and the P3 key is the factory default print key to accomplis...

Страница 153: ...rs related to the service 7 3 3 Setting the P3 Key to USB Quick Save If the P3 Key is not set to USB Quick Save 1 While on the Connectivity screen with the Buttons tab displayed go to the Destinations...

Страница 154: ...Select Connectivity from the Utilities Menu 3 Select the Buttons tab on the Connectivity screen 4 In the Physical Print Buttons field select Print1 The Connectivity Buttons Screen will be displayed l...

Страница 155: ...turing a Screen cont d 3 Click FREEZE to unfreeze the image to view the image screen and the snapshots displayed on the bottom 4 Highlight the snapshot to be stored 5 Select Menu on the right side of...

Страница 156: ...y to Customer s Functionality If the customer had programmed the P3 Key to a function other than screen capture restore that functionality recorded in section 7 4 1 on page 6 Refer to Figure 7 2 1 Cli...

Страница 157: ...7 5 Global Service User Interface GSUI 7 5 1 Enter global service user interface 1 Press Utility Config Service see Figure 7 6 on page 7 9 1 Select GE Service in option of User Level of GEMS Service...

Страница 158: ...agnostic control and Operator feedback 7 5 3 1 Button This button has two modes each with appropriate text Execute to start the diagnostic and Abort to stop a diagnostic The button can also be disable...

Страница 159: ...t Message Displays either a starting message or aborting message as well as the diagnostic completion status 7 5 3 5 Background Color Initially gray the Control Frame background color changes upon com...

Страница 160: ...two selections 7 6 1 1 Disruptive Mode Allows you to enable or disable disruptive mode troubleshooting 7 6 1 2 System Shutdown Allows for system shutdown from the diagnostic menu Select to Restart Sy...

Страница 161: ...agnostics Troubleshooting 7 13 7 6 2 PC Diagnostics Non Interactive Tests 7 6 2 1 Essential Tests 7 6 2 2 System Board 7 6 2 3 Hard Disk Drive Surface Scan 7 6 2 4 Hard Disk Drive Quick Test 7 6 2 5 P...

Страница 162: ...ion 7 6 Common Diagnostics 7 6 3 PC Diagnostics Interactive Tests 7 6 3 1 Keyboard Test 7 6 3 2 Mouse Test 7 6 3 3 Audio Card Test only for LOGIQ e R5 2 x R6 x x R7 x x LOGIQ e Vet R6 x x R7 x x 7 6 4...

Страница 163: ...oubleshooting 7 15 Section 7 7 Network and Insite II Configuration 7 7 1 Network Configuration 7 7 1 1 Wire LAN Network 1 Connect system with network 2 Press Utility Config Connectivity TCPIP in IP Se...

Страница 164: ...setup static IP address uncheck Enable DHCP option input static address in IP Address box Subnet Mask and Default Gateway box In Network Speed box choose the proper speed available 3 Click Save in po...

Страница 165: ...tart at the left bottom of screen and select Network Connections 3 In the pop up window choose Wireless Network Connection Click it with cursor key then select View Available Wireless Networks Figure...

Страница 166: ...OTE If an encrypted network has been selected input Network key in Wireless Network Connection dialog box displayed and click Connect 5 The network icon at the left bottom of screen turns green Figure...

Страница 167: ...page 9 7 To use Pilot Server on Configuration page choose Questra Agent Configuration in left column Select Advanced Configuration then choose Others as server type and manually input pilot address a...

Страница 168: ...site II Configuration cont d 8 Change to Device Information page enter Device and Serial Number as the serial number of that system then press Modify Device Configurations NOTE The Device name and Ser...

Страница 169: ...ivid e Basic Service Manual Chapter 7 Diagnostics Troubleshooting 7 21 7 7 2 Insite II Configuration cont d 9 Change to Proxy_Configuration page enter Proxy Server Address as 3 20 128 6 and Proxy Serv...

Страница 170: ...7 Network and Insite II Configuration 7 7 2 Insite II Configuration cont d 10 Change to Diagnostics page Enable Disruptive Mode at Service side by pressing Yes 11 Customer will receive a request of S...

Страница 171: ...al Chapter 7 Diagnostics Troubleshooting 7 23 7 7 2 Insite II Configuration cont d 12 Service can do Remote Diagnostics successfully 13 Press power switch and choose Full Maintenance Reboot to reboot...

Страница 172: ...R7 x x 1 Enter global service user interface and log in as GE Service See 7 5 1 2 On Configuration page select Insite ExC Agent Configuration in the left column fill in the information in Agent Confi...

Страница 173: ...uptive Mode in the left column Check whether the Disruptive Mode Status is Enabled if not click Yes to change the Disruptive Mode Status as Enabled See Figure 7 28 on page 7 25 There is another way to...

Страница 174: ...3 Insite II Configuration For LOGIQ e R6 x x R7 x x and LOGIQ e Vet R6 x x R7 x x cont d 4 Customer will receive a request of Service side click Yes to accept See Figure 7 30 on page 7 26 5 Service c...

Страница 175: ...7 27 7 7 3 Insite II Configuration For LOGIQ e R6 x x R7 x x and LOGIQ e Vet R6 x x R7 x x cont d 6 Press power switch select Full Maintenance Reboot to reboot system see Figure 7 32 on page 7 27 7 In...

Страница 176: ...3 Insite II Configuration For LOGIQ e R6 x x R7 x x and LOGIQ e Vet R6 x x R7 x x cont d 8 Select Update Software in the category of SERVICE 9 The page of software package shows up select the correct...

Страница 177: ...leshooting 7 29 7 7 3 Insite II Configuration For LOGIQ e R6 x x R7 x x and LOGIQ e Vet R6 x x R7 x x cont d 10 Click icon on the schedule page to set up a schedule Fill in the name and time then clic...

Страница 178: ...con will show in the status bar 13 Press Power button The SYSTEM EXIT window shows up Click Decline the system will not download the software and the Insite icon in the status bar will not shown Click...

Страница 179: ...x R7 x x cont d NOTE Make sure the system administrator is logged on as ADM or the system will indicate such message 14 The system begins to download the software and wait until the process complete...

Страница 180: ...I Configuration 7 7 3 Insite II Configuration For LOGIQ e R6 x x R7 x x and LOGIQ e Vet R6 x x R7 x x cont d 15 When the download is complete it will show the window below 16 Restart the system and th...

Страница 181: ...Vet R6 x x R7 x x cont d 17 After the system is on login as ADM Press Power button and the system exit window displays Select Verify 18 It will show the verify items Select Passed and sign the signatu...

Страница 182: ...GE Direction 5370626 100 Revision 18 LOGIQ e LOGIQ e Vet LOGIQ i Vivid e Basic Service Manual 7 34 Section 7 7 Network and Insite II Configuration This page was intentionally left blank...

Страница 183: ...scription Page Number 8 1 Overview page 8 1 8 2 Disassembly Re assembly of LOGIQ e LOGIQ e Vet LOGIQ i Vivid e page 8 1 8 2 1 Warning and Caution page 8 1 8 2 2 Handle Assy FRU No 313 page 8 2 8 3 Loa...

Страница 184: ...he Handle Assy FRU No 313 8 2 2 1 Tools Plier 8 2 2 2 Needed Manpower 1 person 2 minutes travel 8 2 2 3 Preparations Shut down the system then disconnect the AC DC power cord and remove the battery 8...

Страница 185: ...Direction 5370626 100 Revision 18 LOGIQ e LOGIQ e Vet LOGIQ i Vivid e Basic Service Manual Chapter 8 Replacement Procedures 8 3 8 2 2 Handle Assy FRU No 313 cont d Figure 8 1 Handle Assy Disassembly 1...

Страница 186: ...the system software with DVD Refer to section 8 3 1 Loading Base Image Software with DVD on page 8 5 from page 8 5 to page 8 9 Load the system software with USB memory stick Refer to Refer to section...

Страница 187: ...the disk labeled System Application Software into the DVD ROM drive Connect the DVD ROM drive to the system 2 Properly turn off the scanner by momentarily pressing the Power On Off Switch Select Full...

Страница 188: ...ly the bootable C partition is loaded This option is intended for recovery of a system that will not boot up It is not recommended for application software upgrade because during upgrade process the d...

Страница 189: ...ware with DVD cont d 8 System DVD will be loaded as shown in Figure 8 6 on page 8 7 If the screen displays as show in Figure 8 7 on page 8 7 after input Yes wait about 1 minute until Figure 8 6 on pag...

Страница 190: ...ny key to shut down the system Refer to Figure 8 8 on page 8 8 10 Press any key to reboot system When system first time boot up after upgrading complete you will see the following message 11 Make a re...

Страница 191: ...Procedures 8 9 8 3 1 Loading Base Image Software with DVD cont d NOTE For LOGIQ e R5 x x R6 x x please select a package to run the application NOTE For LOGIQ e R5 x x you need to select a package to r...

Страница 192: ...turn off the scanner by momentarily pressing the Power On Off Switch Select Full Maintenance Reboot Full Reboot from the System Exit menu 2 If the system will not shutdown normally hold down the Powe...

Страница 193: ...e that any patient data on the disk has been backed up To select b only the bootable C partition is loaded This option is intended for recovery of a system that will not boot up It is not recommended...

Страница 194: ...Q e Vet LOGIQ i Vivid e Basic Service Manual 8 12 Section 8 3 Loading the System Software 8 3 2 Loading the System Software with USB Memory Stick cont d 7 Input Yes or No and press Enter key to contin...

Страница 195: ...e 8 16 on page 8 13 Wait for the software installation to complete Typical installation time 5 10 minutes Status bar on the screen indicates progress If the screen displays as show in Figure 8 17 on p...

Страница 196: ...Loading the System Software with USB Memory Stick cont d 9 After finish updating system remove the USB memory stick and press any key to shut down the system Refer to Figure 8 18 on page 8 14 NOTE If...

Страница 197: ...nd select Next to run the application or to abort select shutdown to shut down the system NOTE For LOGIQ e R5 x x and R6 x x please select a package to run the application NOTE When doing an option A...

Страница 198: ...8 3 Loading the System Software 8 3 2 Loading the System Software with USB Memory Stick cont d 11 When select Next in step 10 the system will instruct to reboot the system to active the package as sh...

Страница 199: ...ect preset region 1 Reboot the system 2 Please select Utility System About and then select Additional About Information NOTE After you have selected Utility System About and selected Additional About...

Страница 200: ...tion 8 3 Loading the System Software 8 3 3Reload the Correct Preset Region cont d 4 The system will pop up a message about the selected region preset Check if the selected region preset is correct pre...

Страница 201: ...asic Service Manual Chapter 8 Replacement Procedures 8 19 8 3 3Reload the Correct Preset Region cont d 6 After the system reboots select Utility System About and select Additional About Information to...

Страница 202: ...ensure that the options are activated and working 1 Reboot the system 2 Select Utility Admin System Admin 3 Ensure that all the installed Option keys are displayed and the status of Options are valid...

Страница 203: ...ion is displayed on the right top location of the screen About the probe specification for intended use on LOGIQ e LOGIQ e Vet LOGIQ i Vivid e please refer to 3 6 6 on page 3 28 Plug in at least one o...

Страница 204: ...DICOM Devices Reinstall any DICOM devices used by the customers and check to ensure these DICOM devices work properly The instruction about installing DICOM devices is not incorporated in this manual...

Страница 205: ...unction check 1 The system can be carried with the handle 2 The handle can rotate from 90 to 110 degree Section 8 3 Loading Base Image Software 4 3 1 4 3 2 4 3 7 4 3 8 4 3 9 4 3 10 4 4 8 3 3 8 3 4 8 3...

Страница 206: ...GE Direction 5370626 100 Revision 18 LOGIQ e LOGIQ e Vet LOGIQ i Vivid e Basic Service Manual 8 24 Section 8 4 Checks after FRU replacement Debrief Guidelines This page was intentionally left blank...

Страница 207: ...placement part available from the manufacturer lead time involved KBD Keyboard LCD Liquid Crystal Display BnV Brightness and Volume RX64 Dragon Front Processor Board TMST Master Board TX64 Dragon Tran...

Страница 208: ...r Part Name Replaced By Quant ity FR U LOGIQ e Vivid e LOGIQ i LOGIQ e Vet 001 5196783 ACDC adapter with clamp filter 5196783 3 1 1 x x x x 001A 5196783 2 ACDC adapter with clamp filter 5196783 3 1 1...

Страница 209: ...5370626 100 Revision 18 LOGIQ e LOGIQ e Vet LOGIQ i Vivid e Basic Service Manual Chapter 9 Renewal Parts 9 3 Section 9 4 Operator Console Assy Figure 9 1 OPERATOR CONSOLE ASSY LCD Assy Keyboard Assy...

Страница 210: ...GE Direction 5370626 100 Revision 18 LOGIQ e LOGIQ e Vet LOGIQ i Vivid e Basic Service Manual 9 4 Section 9 5 LCD Assy Section 9 5 LCD Assy Figure 9 2 LCD Assy 100 101 102 103 104 105...

Страница 211: ...x Vivid e R6 x x 5419295 2 1 1 x 100A 5419295 2 LCD Assy for LOGIQ e R7 x x Vivid e R6 x x LOGIQ e Vet R7 x x 1 1 x 101 5133509 3 LCD Front Cover for LOGIQ e R4 x x LOGIQ e R5 0 x LOGIQ e R5 2 x LOGIQ...

Страница 212: ...x x Vivid e R4 x x LOGIQ i R4 x x 5439794 1 2 x x x 105 5125944 3 LCD cable 5125944 4 1 2 x x x x 105A 5125944 4 LCD Cable with JAE connector 1 2 x x x x 5125944 6 6 bits LCD cable for Service LOGIQ e...

Страница 213: ...GE Direction 5370626 100 Revision 18 LOGIQ e LOGIQ e Vet LOGIQ i Vivid e Basic Service Manual Chapter 9 Renewal Parts 9 7 Section 9 6Keyboard Assy Figure 9 3 Keyboard Assy 200 201 202 203...

Страница 214: ...Direction 5370626 100 Revision 18 LOGIQ e LOGIQ e Vet LOGIQ i Vivid e Basic Service Manual 9 8 Section 9 6 Keyboard Assy Section 9 6 Keyboard Assy cont d Figure 9 4 Keyboard Assy cont d 205 204 206 2...

Страница 215: ...458515 and 5439794 1 1 x 5172817 Keyboard Assy for Vivid e R4 x x 5458515 and 5458755 1 1 x 5199342 Keyboard Assy Vivid e R5 x x Vivid e R6 x x 5458515 and 5458755 1 1 x 5184814 Keyboard Assy for LOGI...

Страница 216: ...x x Vivid e R4 x x LOGIQ i R4 x x 5154621 2 1 1 x x x 204A 5154621 2 KBD cable kits for LOGIQ e R4 x x LOGIQ e R5 0 x LOGIQ e R5 2 x LOGIQ e R6 x x LOGIQ e Vet Vivid e R5 x x Vivid e R6 x x LOGIQ i R...

Страница 217: ...ion 5370626 100 Revision 18 LOGIQ e LOGIQ e Vet LOGIQ i Vivid e Basic Service Manual Chapter 9 Renewal Parts 9 11 Section 9 7Bottom Assy Figure 9 5 Bottom Assy 300 301 302 303 304 310 312 301 303 309...

Страница 218: ...irection 5370626 100 Revision 18 LOGIQ e LOGIQ e Vet LOGIQ i Vivid e Basic Service Manual 9 12 Section 9 7 Bottom Assy Section 9 7 Bottom Assy cont d Figure 9 6 Bottom Assy cont d 307 308 310 309 306...

Страница 219: ...ection 5370626 100 Revision 18 LOGIQ e LOGIQ e Vet LOGIQ i Vivid e Basic Service Manual Chapter 9 Renewal Parts 9 13 Section 9 7 Bottom Assy cont d Figure 9 7 Bottom Assy cont d 313 317 316 311 312 31...

Страница 220: ...0 5 1 1 x x 301A 2406640 3R TMST PWA for LOGIQ e R4 x x LOGIQ e R5 x x LOGIQ e R6 x x Vivid e R4 x x Vivid e R5 x x LOGIQ i R4 x x LOGIQ i R5 x x LOGIQ e Vet R5 x x LOGIQ e Vet R6 x x 1 1 x x x x 302...

Страница 221: ...for LOGIQ e R5 0 x LOGIQ e R5 2 x LOGIQ e R6 x x LOGIQ e Vet 5428625 1 1 x x 5148773 Menu Panel Assy for LOGIQ e R4 x x Vivid e R4 x x 5432049 1 1 x x 5212340 Menu Panel Assy for Vivid e R5 x x Vivid...

Страница 222: ...AN Assy 2 Fans 1 1 x x x x 311 5155166 LCD and Handle Hinge kits 1 1 x x x x 312 2406733 TMST2TX64 Assy 2406733 R 1 2 x x x x 312A 2406733 R TMST2TX64 Assy 1 2 x x x x 313 5154937 Handle Assy for LOGI...

Страница 223: ...R7 0 3 or below Vivid e R4 0 x Vivid e R5 0 x Vivid e R6 0 0 Vivid e R6 0 1 LOGIQ e Vet R5 0 x LOGIQ e Vet R6 0 5 or below LOGIQ i R4 1 x LOGIQ i R5 1 x 1 1 x x x x 5478480 SATA HDD with Black Front...

Страница 224: ...x x Vivid e R5 x x Vivid e R6 x x LOGIQ e Vet R6 0 5 or above LOGIQ e Vet R7 x x 5422180 Battery Clip for LOGIQ e R7 x x Vivid e R5 x x Vivid e R6 x x LOGIQ e Vet R6 0 5 or above LOGIQ e Vet R7 x x 5...

Страница 225: ...Bottom Cover Assy for LOGIQ e R4 x x LOGIQ e R5 0 x LOGIQ e R5 2 x LOGIQ e R6 x x LOGIQ e Vet 5135311 Battery Clip for LOGIQ e R5 0 x LOGIQ e R5 2 x LOGIQ e R6 x x LOGIQ e Vet Vivid e R5 x x Vivid e...

Страница 226: ...GE Direction 5370626 100 Revision 18 LOGIQ e LOGIQ e Vet LOGIQ i Vivid e Basic Service Manual 9 20 Section 9 9 Cables Section 9 9 Cables...

Страница 227: ...em Part Number Part Name Replaced By Quantity FRU LOGIQ e LOGIQ e Vet Vivid e LOGIQ i 501 5183729 Wheels 1 2 x x x x 502 5180439 2 Isolation transformer 110V 1 2 x x x x 5180376 Isolation transformer...

Страница 228: ...India South Africa class cable kits 1 2 x x x x 5182881 Argentina class cable kits 1 2 x x x x 5182625 Israel class cable kits 1 2 x x x x 5182631 Switzerland class cable kits 1 2 x x x x 507 5183906...

Страница 229: ...OGIQ e LOGIQ e Vet LOGIQ i Vivid e Basic Service Manual Chapter 9 Renewal Parts 9 23 Section 9 11 Isolation Cart Enhanced Version Components Figure 9 11 Isolation Cart Components 515 511 513 514 516 5...

Страница 230: ...transformer 220V Onyx Black 1 2 x x x x 514 5394028 Cable hook kit Dark Steel Blue 5426642 1 2 x x x x 514A 5426642 Cable hook kit Onyx Black 1 2 x x x x 515 5394065 Isolation Cart Rear Handle Dark S...

Страница 231: ...146 AC Power Cord USA 5177146 2 1 2 x x x x 524A 5177154 2 AC Power Cord Switzerland 1 2 x x x x 5176753 2 AC Power Cord Israel 1 2 x x x x 5177195 2 AC Power Cord Argentina 1 2 x x x x 5176773 2 AC P...

Страница 232: ...x x 602 5151233 DVD RW USA LG DVD RW for LOGIQ e R4 x x Vivid e R4 x x LOGIQ i R4 x x 5449275 2 1 2 x x x 602A 5151233 2 DVD RW USA LITEON DVD RW 5449275 2 1 2 x x x x 602B 5151233 3 DVD RW USA LITEON...

Страница 233: ...7 Digital B W Printer EU 5151261 2 1 2 x x x x 610B 5151261 2 SONY UP D898MD BW Printer Europe Kit 1 2 x x x x 611 5420901 MITSUBISHI P95D BW Printer CHN kit 5151262 1 2 x x x x 611A 5151262 UP D897 D...

Страница 234: ...1 2 x x x 622A 5175833 2 HP K5400 PC Printer EU 5175833 3 1 2 x x x x 622B 5175833 3 HP K8600 PC Printer EU 1 2 x x x x 623 5175554 HP K550 PC Printer CN for LOGIQ e R4 x x Vivid e R4 x x LOGIQ i R4...

Страница 235: ...it for LOGIQ e Vet 1 2 x 5183684 LOGO Kit for Vivid e R4 x x 1 2 x 5183684 LOGO Kit for Vivid e R5 x x Vivid e R6 x x 1 2 x 5191678 LOGO Kit for LOGIQ i R4 x x LOGIQ i R5 x x 1 2 x 635 5173797 USB Har...

Страница 236: ...5421870 1 2 x x x x 644A 5264324 2 USB Wireless adapter WN111v2 Installation kit 5421870 1 2 x x x x 644B 5421870 NetGear N300 Wireless USB Adapter WNA3100 kit 1 2 x x x x 645 5473920 NetGear WNA1000...

Страница 237: ...1 2 x 5396928 4 System and Application software DVD for LOGIQ e R 6 0 2 1 2 x 5396928 5 System and Application software DVD for LOGIQ e R 6 0 3 5396928 6 1 2 x 5447434 System and Application software...

Страница 238: ...plication software USB for LOGIQ e Vet R6 0 5 1 2 x 5459106 2 System and Application software USB for LOGIQ e Vet R6 0 6 1 2 x 5454607 System and Application software USB for LOGIQ e Vet R7 0 0 1 2 x...

Страница 239: ...software DVD for LOGIQ i R5 1 1 1 2 x 655 5443822 Patch CD Kits for LOGIQ e Vivid e and LOGIQ e Vet patch installation purpose only 1 2 x x x 656 5461837 Software Service kit LOGIQ e R6 0 4 software...

Страница 240: ...Installation Kit 1 2 x x x x 673 5499686 eSmartTrainer USB Flash Drive for Japan 1 2 x 674 5500443 Virtual Memory Patch Installation Kit for LOGIQ e R7 x x 1 2 x 675 5495509 SONY UPD897 BW Printer Bra...

Страница 241: ...74 100 Quick Start Guide 1 N LOGIQ e R4 x x Quick Start Guide French 5130174 101 Quick Start Guide 1 N LOGIQ e R4 x x Quick Start Guide Spanish 5130174 106 Quick Start Guide 1 N LOGIQ e R4 x x Quick S...

Страница 242: ...ser Manual Chinese 5198506 141 Basic User Manual for R5 0 x 5 2 x 6 x x 7 x x 1 N LOGIQ e R5 0 x CKD User Manual English 5268200 100 Basic User Manual for R5 0 x 1 N LOGIQ e User Manual English 531462...

Страница 243: ...id e R4 x x Item Part Name Part Number Description Quantity FRU LOGIQ e LOGIQ e Vet LOGIQ i Vivid e Service Manual 5370626 100 Service Manual 1 N System User Manuals Vivid e R4 x x User Manual English...

Страница 244: ...74 111 Basic User Manual 1 N Vivid e R5 x x R6 x x User Manual Portuguese 5198874 127 Basic User Manual 1 N Vivid e R5 x x R6 x x User Manual Japanese 5198874 140 Basic User Manual 1 N Vivid e R5 x x...

Страница 245: ...al 1 N LOGIQ i R5 x x User Manual German 5212229 108 Basic User Manual 1 N LOGIQ i R5 x x User Manual Italian 5212229 111 Basic User Manual 1 N LOGIQ i R5 x x User Manual Portuguese 5212229 127 Basic...

Страница 246: ...er Frequency 6 5 20 MHz 1 1 x x x x x x x x x 703 8C RS 2354971 Probe Center Frequency 6 5 20 MHz 5499508 1 1 x x x x x x x x x x 703A 8C RS 5499508 Probe Center Frequency 6 5 20 MHz 1 1 x x x x x x x...

Страница 247: ...quency 4 85 20 MHz 1 1 x x x x x 713 16L RS 5317271 Probe Center Frequency 12 0 20 MHz 1 1 x x 714 6Tc RS KN10010 4 Probe Center Frequency 4 8 20 MHz x x x 715 L8 18i RS 5397810 Probe Center Frequency...

Страница 248: ...0 MHz 1 1 x x x x 753 i12L RS Vet 5134645 Probe Center Frequency 5 6 20 MHz 1 1 x x x x 754 8L RS Vet 2376127 Probe Center Frequency 6 3 20 MHz 1 1 x x x x 755 3S RS Vet 5134647 Probe Center Frequency...

Страница 249: ...lity of the ultrasound systems performance Read this chapter completely and familiarize yourself with the procedures before performing a task Table 10 1 Contents in Chapter 10 Section Description Page...

Страница 250: ...ditation from organizations such as the American College of Radiology USA it is the customer s responsibility to have a quality assurance program in place for each scanner The program must be directed...

Страница 251: ...es and recommended intervals shown in the Care Maintenance Task Schedule assumes that you use your LOGIQ e LOGIQ e Vet LOGIQ i Vivid e for an average patient load 10 12 per day and use it as a primary...

Страница 252: ...6 194427P373 46 194427P370 Kit includes anti static mat wrist strap and cables for 200 to 240 V system 3M 2204 Large adjustable wrist strap 3M 2214 Small adjustable wrist strap 3M 3051 conductive grou...

Страница 253: ...have any problems or questions about the equipment 2 Paperwork Fill in the top of the Ultrasound Inspection Certificate see page 19 Note all probes and system options 3 Power up With AC input Turn the...

Страница 254: ...s mode of operation M Mode Verify basic M Mode operation Check the basic system controls that affect this mode of operation Applicable Software Options Verify the basic operation of all optional modes...

Страница 255: ...is functioning as programed Clean as necessary 5 ECG Verify basic operation with customer 6 DVD Verify that the DVD is functioning properly Clean heads and covers if necessary Table 10 7 AC DC Adapte...

Страница 256: ...Control Panel Inspect keyboard and control panel Note any damaged or missing items 4 Cables Connectors Check all internal cable harnesses and connectors for wear and secure connector seating Pay speci...

Страница 257: ...equipment damage Failure to properly handle or maintain a probe may also void its warranty Any evidence of wear indicates the probe cannot be used Do a visual check of the probe pins and system socke...

Страница 258: ...year Please follow the flow chart below to carry out battery performance maintenance NOTE Disconnect all probes when discharge battery NOTE Discharge the battery to let the system automatically shut d...

Страница 259: ...current To minimize the risk that a probe may shock someone the customer should Not use a probe that is cracked or damaged in any way Check probe leakage current Based on your facilities QA program f...

Страница 260: ...e site These tests are chassis leakage current and probe leakage current All measurements are made with an electrical safety analyzer which should be calibrated and compliant with AAMI ESI 1993 or IEC...

Страница 261: ...the combination of lights that are illuminated Any problems found should be reported to the hospital immediately and the receptacle should not be used NOTE No outlet tester can detect the condition w...

Страница 262: ...e made with the unit ON and OFF with the power line polarity Normal and Reversed Record the highest reading of current When using the Microguard or a similar test instrument its power plug may be inse...

Страница 263: ...value shown in Table 10 11 Record all data on the PM Inspection Certificate Table 10 14 Typical Data Sheet for enclosure Source Leakage Current Unit Power Tester Polarity Switch Tester Neutral or Grou...

Страница 264: ...On For each combination the probe must be active to find the worst case condition NOTE Each probe will have some amount of leakage current dependent on its design Small variations in probe leakage cu...

Страница 265: ...pment damage possibility Never switch the Polarity and the status of Neutral when the unit is powered ON Be sure to turn the unit power OFF before switching them using the POLARITY switch and or the N...

Страница 266: ...E FAILS Change another probe to confirm if the fail is caused by console NOTE Each probe will have some amount of leakage dependent on its design Small variations in probe leakage currents are normal...

Страница 267: ...Model Number Serial Number Manufacture Date Probe 1 Frequency Scan Format Model Number Serial Number Probe 2 Frequency Scan Format Model Number Serial Number Probe 3 Frequency Scan Format Model Number...

Страница 268: ...system covers are in place System scans with all probes as expected Accepted by ______________________________________________________________________ Electrical Test Performed Max Value Allowed Valu...

Страница 269: ...Procedure Warnings 1 15 Date Time display location 4 9 Depth scale display location 4 9 Diagnostics 5 17 DICOM Network Function 2 9 E Electrical requirements 2 2 Electrical Safety 1 13 Electrostatic...

Страница 270: ...e Tests 7 14 Peripherals Functional Checks 4 37 Power On 3 9 Power Requirements 2 2 electrical 2 2 stability 2 3 Power Stability Requirements 2 3 Probe Connector Cleaning 10 9 Probe identifier display...

Страница 271: ......

Страница 272: ...2004 to 2016 General Electric Company GE Medical Systems a General Electric Company going to market as GE Healthcare 3200 N Grandview Boulevard Waukesha Wisconsin 53149 USA www gehealthcare com...