Probes
265
LOGIQ
t
α
100 User Manual
2211157–100 Rev 0
Disinfecting probes
Ultrasound probes can be disinfected using liquid
chemical germicides. The level of disinfection is directly
related to the duration of contact with the germicide.
Increased contact time produces a higher level of
disinfection. 2% Glutaraldehyde-based solutions have
been shown to be very effective for this purpose. Cidex is
the only germicide that has been evaluated compatibility
with the material use to construct the probes.
In order for liquid chemical germicides to be effective, all
visible residue must be removed during the cleaning
process. Thoroughly clean the probe, as described
earlier, before attempting disinfection.
1. Prepare the germicide solution according to the
manufacturer’s instructions. Be sure to follow all
precautions for storage, use and disposal.
2. Place the cleaned and dried probe surface in
contact with the germicide solution. High-level
disinfection (for surface and cavity probes) can be
accomplished by soaking according to the contact
time recommended by the germicide
manufacturer.
3. After removing from the germicide, rinse the
probe following the germicide manufacturer’s
rinsing instructions. Flush germicide residue from
the probe and allow to air dry.
After Each Use
CAUTION
Содержание LOGIQ A100
Страница 18: ...Introduction LOGIQt α100 User Manual 2211157 100 Rev 0 2 This page left blank intentionally ...
Страница 30: ...Getting Started LOGIQt α100 User Manual 2211157 100 Rev 0 14 This page left blank intentionally ...
Страница 42: ...Preparing the System for Use 26 LOGIQt α100 User Manual 2211157 100 Rev 0 This page left blank intentionally ...
Страница 54: ...Safety LOGIQt α100 User Manual 2211157 100 Rev 0 38 This page left blank intentionally ...
Страница 62: ...Patient Safety 46 LOGIQt α100 User Manual 2211157 100 Rev 0 This page left blank intentionally ...
Страница 74: ...Device Labels 58 LOGIQt α100 User Manual 2211157 100 Rev 0 This page left blank intentionally ...
Страница 76: ...Safety LOGIQt α100 User Manual 2211157 100 Rev 0 60 This page left blank intentionally ...
Страница 80: ...Patient Registration 64 LOGIQt α100 User Manual 2211157 100 Rev 0 This page left blank intentionally ...
Страница 88: ...Image Display 72 LOGIQt α100 User Manual 2211157 100 Rev 0 This page left blank intentionally ...
Страница 100: ...VCR Operations 84 LOGIQt α100 User Manual 2211157 100 Rev 0 This page left blank intentionally ...
Страница 114: ...General Measurements LOGIQt α100 User Manual 2211157 100 Rev 0 98 This page left blank intentionally ...
Страница 140: ...Diagnostic Category LOGIQt α100 User Manual 2211157 100 Rev 0 124 This page left blank intentionally ...
Страница 146: ...OB LOGIQt α100 User Manual 2211157 100 Rev 0 130 This page left blank intentionally ...
Страница 148: ...Exam Preparation 132 LOGIQt α100 User Manual 2211157 100 Rev 0 This page left blank intentionally ...
Страница 188: ...OB Measurements 172 LOGIQt α100 User Manual 2211157 100 Rev 0 This page intentionally left blank ...
Страница 224: ...OB Report Page 208 LOGIQt α100 User Manual 2211157 100 Rev 0 This page left blank intentionally ...
Страница 236: ...Cardiology 220 LOGIQt α100 User Manual 2211157 100 Rev 0 This page left blank intentionally ...
Страница 248: ...Urology LOGIQt α100 User Manual 2211157 100 Rev 0 232 This page left blank intentionally ...
Страница 251: ...LOGIQt α100 User Manual 2211157 100 Rev 0 235 Control Keys Configuration Using Control Keys ...
Страница 252: ...Control Keys LOGIQt α100 User Manual 2211157 100 Rev 0 236 This page left blank intentionally ...
Страница 272: ...Configuration Using Control Keys 256 LOGIQt α100 User Manual 2211157 100 Rev 0 This page left blank intentionally ...
Страница 274: ...Probes Biopsy LOGIQt α100 User Manual 2211157 100 Rev 0 258 This page left blank intentionally ...
Страница 298: ...Probes 282 LOGIQt α100 User Manual 2211157 100 Rev 0 This page left blank intentionally ...
Страница 326: ...Troubleshooting and Maintenance LOGIQt α100 User Manual 2211157 100 Rev 0 310 This page left blank intentionally ...
Страница 344: ...Maintenance 328 LOGIQt α100 User Manual 2211157 100 Rev 0 This page left blank intentionally ...
Страница 346: ...OB Tables LOGIQt α100 User Manual 2211157 100 Rev 0 330 This page left blank intentionally ...
Страница 400: ...OB Tables 384 LOGIQt α100 User Manual 2211157 100 Rev 0 This page left blank intentionally ...
Страница 408: ...Index Index 8 LOGIQt α100 User Manual 2211157 100 Rev 0 This page left blank intentionally ...