Probes
266
LOGIQ
t
α
100 User Manual
2211157–100 Rev 0
Disinfecting probes (
cont’d
)
Special Disinfecting Instructions for the E72: To properly
disinfect the E72 probe, the probe handle can be
reassembled loosely so that the entire probe with handle
can be immersed in the germicide solution. The
adjustment screw must be kept loose so that germicide
can penetrate to all surfaces. After immersing, rotate
and shake the probe while it is below the surface of the
germicide to eliminate air pockets. Allow the germicide
to remain in contact with the fully immersed probe, for
high level disinfection, according to the germicide
manufacturer’s recommended time. To remove all
germicide residue, final rinsing should be done following
the germicide manufacturer’s instructions. Remove
excess water by shaking and allow to air dry.
Ultrasound transducers can easily be damaged by
improper handling and by contact with certain chemicals.
Failure to follow these precautions can result in serious
injury and equipment damage.
S
Do not immerse the probe into any liquid beyond the
level specified for that probe. Never immerse the
transducer connector or probe adapters into any
liquid.
S
Avoid mechanical shock or impact to the transducer
and do not apply excessive bending or pulling force
to the cable.
S
Transducer damage can result from contact with
inappropriate coupling or cleaning agents:
S
Do not soak or saturate transducers with solutions
containing alcohol, bleach, ammonium chloride
compounds or hydrogen peroxide
S
Avoid contact with solutions or coupling gels
containing mineral oil or lanolin
S
Avoid temperatures above 60
°
C.
S
Inspect the probe prior to use for damage or
degeneration to the housing, strain relief, lens and
seal. Do not use a damaged or defective probe.
.
WARNING
Содержание LOGIQ A100
Страница 18: ...Introduction LOGIQt α100 User Manual 2211157 100 Rev 0 2 This page left blank intentionally ...
Страница 30: ...Getting Started LOGIQt α100 User Manual 2211157 100 Rev 0 14 This page left blank intentionally ...
Страница 42: ...Preparing the System for Use 26 LOGIQt α100 User Manual 2211157 100 Rev 0 This page left blank intentionally ...
Страница 54: ...Safety LOGIQt α100 User Manual 2211157 100 Rev 0 38 This page left blank intentionally ...
Страница 62: ...Patient Safety 46 LOGIQt α100 User Manual 2211157 100 Rev 0 This page left blank intentionally ...
Страница 74: ...Device Labels 58 LOGIQt α100 User Manual 2211157 100 Rev 0 This page left blank intentionally ...
Страница 76: ...Safety LOGIQt α100 User Manual 2211157 100 Rev 0 60 This page left blank intentionally ...
Страница 80: ...Patient Registration 64 LOGIQt α100 User Manual 2211157 100 Rev 0 This page left blank intentionally ...
Страница 88: ...Image Display 72 LOGIQt α100 User Manual 2211157 100 Rev 0 This page left blank intentionally ...
Страница 100: ...VCR Operations 84 LOGIQt α100 User Manual 2211157 100 Rev 0 This page left blank intentionally ...
Страница 114: ...General Measurements LOGIQt α100 User Manual 2211157 100 Rev 0 98 This page left blank intentionally ...
Страница 140: ...Diagnostic Category LOGIQt α100 User Manual 2211157 100 Rev 0 124 This page left blank intentionally ...
Страница 146: ...OB LOGIQt α100 User Manual 2211157 100 Rev 0 130 This page left blank intentionally ...
Страница 148: ...Exam Preparation 132 LOGIQt α100 User Manual 2211157 100 Rev 0 This page left blank intentionally ...
Страница 188: ...OB Measurements 172 LOGIQt α100 User Manual 2211157 100 Rev 0 This page intentionally left blank ...
Страница 224: ...OB Report Page 208 LOGIQt α100 User Manual 2211157 100 Rev 0 This page left blank intentionally ...
Страница 236: ...Cardiology 220 LOGIQt α100 User Manual 2211157 100 Rev 0 This page left blank intentionally ...
Страница 248: ...Urology LOGIQt α100 User Manual 2211157 100 Rev 0 232 This page left blank intentionally ...
Страница 251: ...LOGIQt α100 User Manual 2211157 100 Rev 0 235 Control Keys Configuration Using Control Keys ...
Страница 252: ...Control Keys LOGIQt α100 User Manual 2211157 100 Rev 0 236 This page left blank intentionally ...
Страница 272: ...Configuration Using Control Keys 256 LOGIQt α100 User Manual 2211157 100 Rev 0 This page left blank intentionally ...
Страница 274: ...Probes Biopsy LOGIQt α100 User Manual 2211157 100 Rev 0 258 This page left blank intentionally ...
Страница 298: ...Probes 282 LOGIQt α100 User Manual 2211157 100 Rev 0 This page left blank intentionally ...
Страница 326: ...Troubleshooting and Maintenance LOGIQt α100 User Manual 2211157 100 Rev 0 310 This page left blank intentionally ...
Страница 344: ...Maintenance 328 LOGIQt α100 User Manual 2211157 100 Rev 0 This page left blank intentionally ...
Страница 346: ...OB Tables LOGIQt α100 User Manual 2211157 100 Rev 0 330 This page left blank intentionally ...
Страница 400: ...OB Tables 384 LOGIQt α100 User Manual 2211157 100 Rev 0 This page left blank intentionally ...
Страница 408: ...Index Index 8 LOGIQt α100 User Manual 2211157 100 Rev 0 This page left blank intentionally ...