Page 3-10
Section 3-5 - Completing the Installation
Installation of Two Probe Adapter & LOGIQ
α
200 Probe Adapter (Optional) (contd.,)
4.)
Ensure that no probe is connected to the system. If connected remove it first.
5.)
Remove the probe holder by lifting it vertically up
6.)
Make sure that the two probe port or LOGIQ
α
200 Probe Adapter lock points to the
12 o'clock
position.
7.)
Align the Two Probe Port or LOGIQ
α
200 Probe Adapter Connector with the System Probe
Port and carefully push it
in.
8.) Align the slots on the upper flange to its counterparts on the top cover of the system.
9.)
Turn the Two Probe Port or LOGIQ a 200 Lock to the 3 o'clock position to secure the Two
Probe Port or LOGIQ a 200 Probe Adapter to
the
system.
10.) Carefully replace the probe holder to its place and press firmly till the projections at
the bottom fit into the slots in the upper flange and the system.
Figure 3-7 Installing Two Probe Port or LOGIQ 200 Probe Adapter
Probe Holder
Two Probe port
Two Probe Port Lock 12 o'clock
Two Probe Port Lock 3 o'clock
Two Probe Port or
LOGIQ a 200 Probe Adapter lock
Hole on the upper flange
(Un Locked Position)
(Locked Position)
LOGIQ a 200 Probe Adapter