Page 2-2
Section 2-2 - General Console Requirements
2-2-1-1
Lighting
Bright light is needed for system installation, updates and repairs. However, operator and patient
comfort may be optimized if the room light is subdued and indirect. Therefore a combination lighting
system (dim/bright) is recommended. Keep in mind that lighting controls and dimmers can be a
source of EMI which could degrade image quality. These controls should be selected to minimize
possible interference.
2-2-2
Electrical Requirements
2-2-2-1
LOGIQ™ 100 PRO Power Requirements
Electrical Specifications for LOGIQ™ 100 PRO.
2-2-2-2
Site Circuit Breaker
It is recommended that the branch circuit breaker for the machine be readily accessible.
CAUTION
POWER OUTAGE MAY OCCUR. The LOGIQ™ 100 PRO requires a dedicated single branch
circuit. To avoid circuit overload and possible loss of critical care equipment, make sure you
DO NOT have any other equipment operating on the same circuit.
2-2-2-3
Site Power Outlets
A dedicated AC power outlet must be within reach of the unit without extension cords. Other outlets
adequate for the external peripherals, medical and test equipment needed to support this unit must
also be present within 1 m (3.2 ft.) of the unit. Electrical installation must meet all current local, state,
and national electrical codes.
2-2-2-4
Unit Power Plug
If the unit arrives without a power plug, or with the wrong plug, you must contact your GE dealer or
the installation engineer must supply what is locally required.
2-2-2-5
Power Stability Requirements
Voltage drop-out
Max 10 ms.
Power Transients
(All applications)
Less than 25% of nominal peak voltage for less than 1 millisecond for any type of transient,
including line frequency, synchronous, asynchronous, or aperiodic transients.
GE Healthcare
Hcat number
Voltage
Tolerances
Current
Frequency
H41072LA,H41072LB,
H41072LC,H41072LD,
H41072LE,H41072LF,
H41072LS
100-230
VAC
±10%
0.9 -1.8 A
Max.
50-60 Hz
Table 2-3 Electrical Specifications for
LOGIQ™ 100 PRO