10
- [EN] English
K335 Issue 3
Calibration - Procedures (Pressure P1/P2)
1. Connect the instrument to the pressure standard
2. Let the equipment get to a stable temperature
(minimum: 30 minutes since the last power on).
3. Use the calibration menu (Table 7) to do a two-point
calibration (Zero and +FS) or a three-point calibration
(-FS, Zero and +FS). Refer to Table 8. The display shows
the applicable instructions to complete the calibration.
Table 8:
Calibration pressures
4. To make sure that the calibration is correct, select the
applicable pressure task (Table 2) and apply these
pressure values:
• Ranges g/d or sg: 0, 20, 40, 60, 80, 100 (%FS)
Then: Go back to 0 in the same steps.
Then (three-point calibration only):
-20, -40, -60, -80, -100 (%FS)
Then: Go back to 0 in the same steps.
• Ranges a: 0, 20, 40, 60, 80, 100 (%FS)
Then: Go back to 0 in the same steps.
5. Make sure that the error is in the specified limits:
• Standard accuracy:
The specified accuracy (refer to “Specification data”)
includes an allowance for temperature changes, reading
stability for one year, and the uncertainty of the standard
used for calibration.
In step 4, make sure that the error between the applied
pressure and the reading on the unit is not more than
0.015% FS.
• Premier precision:
The specified precision (refer to “Specification data”)
includes an allowance for temperature changes and the
uncertainty of the standard used for calibration.
In step 4, make sure that the error between the applied
pressure and the reading on the unit is not more than the
specified value for
Premier precision
.
Calibration - Procedures (mA input)
1. DPI 802/802P only. Connect the instrument to the
calibration equipment (Figure 7).
2. Let the equipment get to a stable temperature
(minimum: 5 minutes since the last power on).
3. Use the calibration menu (Table 7) to do a three-point
calibration (-FS, Zero and +FS). The display shows the
applicable instructions to complete the calibration.
4. To make sure that the calibration is correct, select the
applicable mA task (Table 2) and apply these values:
• mA: -55, -40, -24, -18, -12, -6, 0 (short circuit)
Then: 0, 6, 12, 18, 24, 40, 55.
5. Make sure that the error is in the specified limits
Table 9:
mA input error limits
Calibration - Procedures (IDOS UMM)
Refer to the user manual for the IDOS UMM.
When the calibration is complete, the instrument
automatically sets a new calibration date in the UMM.
Ranges: g/d
Nominal applied pressure
psi (mbar)
-FS †
Zero
+FS
≤
10.0 psi (700 mbar)
-FS
0
+FS
> 10.0 psi (700 mbar)
-13.1 (-900)
0
+FS
†
For a three-point calibration, do not apply more than -90% of
the specified FS for the unit.
Ranges: a
Nominal applied pressure
psi (mbar)
Zero
+FS
5.00 psi (350 mbar)
< 0.02 (1.0)
+FS
30.0 psi (2 bar)
< 0.07 (5.0)
+FS
100.0 psi (7 bar)
< 0.29 (20.0)
+FS
300.0 psi (20 bar)
< 0.73 (50.0)
+FS
Ranges: sg
Nominal applied pressure
Zero
+FS
≥
5000 psi (350 bar)
Use atmospheric
pressure as zero.
+FS
Applied
mA
Calibrator
error
(mA)
Permitted
DPI 800/802 error
(mA)
±55
0.0022
0.005
±40
0.0018
0.004
±24
0.0014
0.003
±18
0.0004
0.003
±12
0.0003
0.002
±6
0.0002
0.002
0 (Short circuit)
-
0.001