background image

2

 - [EN]  English

K335 Issue 3

Safety

Before you use the instrument, make sure that you read 
and understand all the related data. This includes: all local 
safety procedures, the instructions for the UMM (if 
applicable), and this publication.

WARNING

• Some liquid and gas mixtures are dangerous. This 

includes mixtures that occur because of 
contamination. Make sure that the instrument is 
safe to use with the necessary media.

• It is dangerous to ignore the specified limits for the 

instrument or to use the instrument when it is not in 
its normal condition. Use the applicable protection 
and obey all safety precautions. 

• To prevent a dangerous release of pressure, isolate 

and bleed the system before you disconnect a 
pressure connection.

• Do not use the instrument in locations with explosive 

gas, vapor or dust. There is a risk of an explosion.

• DPI 802/802P only. To prevent electrical shocks or 

damage to the instrument, do not connect more 
than 30V between the terminals, or between the 
terminals and the ground (earth).

Before you start an operation or procedure in this 
publication, make sure that you have the necessary skills 
(if necessary, with qualifications from an approved training 
establishment). Follow good engineering practice at all 
times.

Safety - Marks and symbols on the instrument

The identification on each pressure connector includes the 
type of thread (1/8 NPT, G1/8 ... ).

To start

To start - Location of items 

 

 …

 

Complies with European 
Union directives

Warning - refer to the 
manual

Read the manual

Battery

Ground (Earth)

 P1/P2 Identifies the pressure 

range connector

ON/OFF

MWP

The maximum working pressure for the connector. 
Example: 2 x FS (FS = Full scale pressure)

REFERENCE

The connector is a reference port for reference pressure only. Refer 
to “Specification data”.

Item

Description

1.

On or off button. 

2.

 

Left-hand soft-key. Selects the function above it on 
the display (Item 24). Example: Edit

3.

ESC

Moves back one menu level.
Leaves a menu option.
Cancels the changes to a value.

4.

Increases or decreases a value.
Highlights a different item.

5.

HOLD

Holds the data on the display. To continue, press the 

HOLD

 button again. 

6.

MENU

OK

Shows the 

Select Task

 menu.

Selects or accepts an item or value.
Selects [

] or cancels [  ] a selection 

7.

 

Right-hand soft-key. Selects the function above it on 
the display (Item 24). Example: Settings

8.

Display. Refer to A3

9.

SENSOR/ 

PC

Communications port. Use to connect a Universal 
Measurement Module (UMM) or a RS232 cable.

10.

10a.

IDOS pressure connector with identification marks. 
Refer to “Safety”

10b.

Label to identify the pressure range for the 
associated IDOS pressure connector (P1/P2): Sealed 
gage (sg), gage (g), absolute (a).

10c.

PTFE vent filter or a reference port connector. Refer to 
“Specification data”.

11.

Connection point for some of the optional 
accessories. Refer to the datasheet.

12.

Battery compartment. Refer to B1.

13., 14., 15.

DPI 802/802P only. Terminals to measure current, to 
supply 24V source, and to do switch tests.

A1

A2

Содержание Druck DPI 800

Страница 1: ...Druck DPI 800 802 Pressure indicator and Pressure loop calibrator User manual K335 GE Sensing Advanced Test Equipment Rentals www atecorp com 800 404 ATEC 2832 Established 1981 ...

Страница 2: ...K335 Issue 3 A1 A2 DPI 802 802P A3 10 a b c 1 3 4 2 5 6 7 9 8 A 2 A 3 11 B1 12 10b 10a 10c 14 15 13 20 21 22 24 17 18 25 19 26 16 23 B1 Customer service Visit our web site www gesensing com ...

Страница 3: ...rademarks All product names are trademarks of their respective companies Introduction The DPI 800 Pressure Indicator and DPI 802 Pressure Loop Calibrator are part of the Druck DPI 800 series of hand held instruments The DPI 800 series uses Intelligent Digital Output Sensor IDOS technology to give instant plug and play functionality with a range of Universal Measurement Modules UMM Example the Univ...

Страница 4: ... G1 8 To start To start Location of items Complies with European Union directives Warning refer to the manual Read the manual Battery Ground Earth P1 P2 Identifies the pressure range connector ON OFF MWP The maximum working pressure for the connector Example 2 x FS FS Full scale pressure REFERENCE The connector is a reference port for reference pressure only Refer to Specification data Item Descri...

Страница 5: ...tifies the type of data and the measurement range Input IDOS input Refer to Select Task Table 2 3 22 23 Identifies the settings applied to the input Refer to Settings Table 4 22 xxxx g The units and the type of sensor If applicable OR a specified scale x y 23 Shows the settings applied to the measured value If applicable Filter Tare Maximum Average Minimum 24 A soft key function To select an avail...

Страница 6: ... to measure the pressure difference between P1 and P2 or P1 P2 and a UPM P High range IDOS Low range IDOS mA DPI 802 802P only A mA measurement task mA 24V DPI 802 802P only A mA measurement task the loop power supply is on DPI 802 802P only A switch test IDOS UMM only An IDOS measurement task Pressure options only A leak test To set up the way the instrument works Additional data Refer to Set Up ...

Страница 7: ...he measurement task Band as a of full scale FS The filter compares each new value with the previous value If the new value is outside the band it is not filtered Low pass filter time constant in seconds Increase the value to increase the damping factor Additional data Select Settings To select and set up the alarm values for the measurement task maximum and minimum Additional data Select Settings ...

Страница 8: ...system Figure 2 Example configuration Measure pressure The examples in Figure 2 show how to measure one or two pressures in a system how to measure pressure difference Note P High range IDOS Low range IDOS If Pressure p2 Pressure p1 the result is negative Operation Leak test To do a leak test on a pressure system 1 Connect the instrument Figure 3 and if necessary adjust the Set Up Table 1 2 Select...

Страница 9: ...e test again press ESC to reset the values 7 Before you disconnect the instrument isolate and bleed the system Figure 4 Example configuration Switch test Operation Transmitter calibration DPI 802 802P only To calibrate a transmitter 1 Connect the instrument Figure 5 or Figure 6 and if necessary adjust the Set Up Table 1 2 Select the applicable calibration task from Select Task Table 2 3 and if nec...

Страница 10: ...Figure 8 Figure 8 Example configuration Pressure measurement with a UPM When the connections are complete make the necessary IDOS selections Table 2 3 Then use the same procedures as specified for the P1 P2 connections If you re attach a UPM the instrument uses the same measurement units that you used before The instrument keeps a record for the last 10 modules Operation Error indications If the d...

Страница 11: ...ent 70 2 F 21 1 C Table 6 Calibration equipment Before you start the calibration make sure that the time and date on the instrument are correct Table 1 Selection sequence Select Task Table 2 Set Up Table 1 Calibration Table 7 Calibration options When you select a channel the display shows the applicable instructions to complete the calibration When the calibration is complete select Calibration Du...

Страница 12: ...sure that the error between the applied pressure and the reading on the unit is not more than the specified value for Premier precision Calibration Procedures mA input 1 DPI 802 802P only Connect the instrument to the calibration equipment Figure 7 2 Let the equipment get to a stable temperature minimum 5 minutes since the last power on 3 Use the calibration menu Table 7 to do a three point calibr...

Страница 13: ...urs P1 mA 5 hours 24 V Source at 12 mA Ranges Gage and differential operation g d sealed gage sg absolute a Type Standard Accuracy FS Premier Precision FS Notes psi 0 36 mbar 25 g d 0 1 0 03 1 2 psi 1 3 5 10 mbar 70 200 350 700 g d 0 075 0 03 1 2 psi 15 to 15 or 30 bar 1 to 1 or 2 g d 0 05 0 01 1 2 psi 15 to 50 100 150 or 300 bar 1 to 3 5 7 10 or 20 g d 0 05 0 01 1 3 psi 500 1000 1500 2000 3000 ba...

Страница 14: ... psi g 2 bar g Premier only 1 8 NPT female port 10 32 UNF reference port OR G1 8 female port M5 reference port Two 1 8 NPT female port 10 32 UNF reference port OR Two G1 8 female port M5 reference port sg or a all ranges G1 8 female port OR 1 8 NPT female port Two 1 8 NPT female port OR Two G1 8 female port Ranges g d sg a MWP Maximum transient intermittent pressure 5 psi 350 mbar 2 x FS 4 x FS 5 ...

Страница 15: ...Customer Service Visit our web site www gesensing com ...

Отзывы: