7.3 Measurement mode TopCOAT
Applications
The TopCOAT method is used for the determination of
wall thickness and coating thickness on coated test ob-
jects.
It is thus possible to carry out residual wall thickness
measurements through coatings on the basis of the first
backwall echo.
The sound velocities in the test object and in the coating
must be known for the application of the measurement
mode TopCOAT. These sound velocities can be
●
entered manually,
●
determined using the measurement mode Auto-V
(see Chapter 7.2
page 7-6),
●
determined by means of calibration blocks (described
later in this section).
Choosing the probe
To be able to use the TopCOAT function, you have to
connect a special 4-element probe to the DMS Go+.
The following probes from GE are suitable for the Top-
COAT method:
●
TC-560
Activating TopCOAT
– Connect a suitable probe to the instrument.
– In the function group
PROBE&CAL
, select the func-
tion
PROBE
.
– Choose the connected probe. The instrument is im-
mediately adjusted.
– Select the function
TG MODE
and choose the option
TOP-COAT
.
Содержание DMS Go+
Страница 1: ...DMS Go Technical Reference and Operating Manual ...
Страница 3: ...DMS Go Edition 4 05 2014 0 3 Function groups and functions ...
Страница 4: ...0 4 Edition 4 05 2014 DMS Go Function groups and functions cont d ...
Страница 8: ...0 8 Edition 4 05 2014 DMS Go ...
Страница 17: ...DMS Go Edition 4 05 2014 1 1 Introduction 1 ...
Страница 41: ...DMS Go Edition 4 05 2014 2 1 Standard package and accessories 2 ...
Страница 54: ...2 Standard package and accessories Recommended accessories 2 14 Edition 4 05 2014 DMS Go ...
Страница 55: ...DMS Go Edition 4 05 2014 3 1 Initial start up 3 ...
Страница 67: ...DMS Go Edition 4 05 2014 4 1 Principles of operation 4 ...
Страница 78: ...4 12 Edition 4 05 2014 DMS Go ...
Страница 79: ...DMS Go Edition 4 05 2014 5 1 Operation 5 ...
Страница 133: ...DMS Go Edition 4 05 2014 6 1 Data Recorder 6 ...
Страница 174: ...6 Data Recorder Documenting measurements 6 42 Edition 4 05 2014 DMS Go ...
Страница 175: ...DMS Go Edition 4 05 2014 7 1 Special functions of the DMS Go TC 7 ...
Страница 189: ...DMS Go Edition 4 05 2014 8 1 Maintenance and care 8 ...
Страница 195: ...DMS Go Edition 4 05 2014 9 1 Interfaces and Peripherals 9 ...
Страница 199: ...DMS Go Edition 4 05 2014 10 1 Appendix 10 ...
Страница 227: ...Recycling directives 10 Appendix DMS Go Edition 4 05 2014 10 29 10 9 Recycling directives Overview 2 3 3 ...
Страница 228: ...10 Appendix Recycling directives 10 30 Edition 4 05 2014 DMS Go 4 5 1 4 ...
Страница 230: ...10 Appendix Recycling directives 10 32 Edition 4 05 2014 DMS Go Materials to be disposed of separately 1 ...
Страница 231: ...Recycling directives 10 Appendix DMS Go Edition 4 05 2014 10 33 2 ...
Страница 233: ...Recycling directives 10 Appendix DMS Go Edition 4 05 2014 10 35 Other materials and components 2 3 3 ...
Страница 234: ...10 Appendix Recycling directives 10 36 Edition 4 05 2014 DMS Go 1 3 ...
Страница 235: ...Recycling directives 10 Appendix DMS Go Edition 4 05 2014 10 37 1 1 ...
Страница 236: ...10 Appendix Recycling directives 10 38 Edition 4 05 2014 DMS Go 5 ...
Страница 237: ...Recycling directives 10 Appendix DMS Go Edition 4 05 2014 10 39 4 7 ...
Страница 238: ...10 Appendix Recycling directives 10 40 Edition 4 05 2014 DMS Go 3 6 6 6 7 ...
Страница 242: ...10 Appendix Recycling directives 10 44 Edition 4 05 2014 DMS Go ...
Страница 243: ...DMS Go Edition 4 05 2014 11 1 Specifications 11 ...
Страница 252: ...11 10 Edition 4 05 2014 DMS Go ...
Страница 253: ...DMS Go Edition 4 05 2014 12 1 Index 12 ...