Carrying out the calibration
5 Operation
DMS Go+
Edition 4 (05/2014)
5-27
– Press one of the
left
or
right
arrow keys in order to
start the zeroing. The message
Wipe off couplant
from probe, press ZERO
is displayed on the screen.
– Press the function key 2 (function
ZERO
). The zero-
ing is performed.
Calibration to zero and to sound velocity
To calibrate the DMS Go+ to the sound velocity in the
test object, you can choose between two methods, i.e.
the 1-point calibration and the 2-point calibration.
As an alternative, if the sound velocity is known, you can
set it directly on the instrument (see Section
The 1-point calibration is available for all measurement
modes, the 2-point calibration only for the measurement
modes S-IP, S-PEAK, S-FLANK, and DUAL.
Zeroing is always carried out first during the 1-point cal-
ibration.
ATTENTION
The reference block at the instrument rear is
only used for carrying out the zeroing pro-
cess.
Only use the calibration blocks recommend-
ed by GE for further calibrations (please refer
to chapter 2).
Содержание DMS Go+
Страница 1: ...DMS Go Technical Reference and Operating Manual ...
Страница 3: ...DMS Go Edition 4 05 2014 0 3 Function groups and functions ...
Страница 4: ...0 4 Edition 4 05 2014 DMS Go Function groups and functions cont d ...
Страница 8: ...0 8 Edition 4 05 2014 DMS Go ...
Страница 17: ...DMS Go Edition 4 05 2014 1 1 Introduction 1 ...
Страница 41: ...DMS Go Edition 4 05 2014 2 1 Standard package and accessories 2 ...
Страница 54: ...2 Standard package and accessories Recommended accessories 2 14 Edition 4 05 2014 DMS Go ...
Страница 55: ...DMS Go Edition 4 05 2014 3 1 Initial start up 3 ...
Страница 67: ...DMS Go Edition 4 05 2014 4 1 Principles of operation 4 ...
Страница 78: ...4 12 Edition 4 05 2014 DMS Go ...
Страница 79: ...DMS Go Edition 4 05 2014 5 1 Operation 5 ...
Страница 133: ...DMS Go Edition 4 05 2014 6 1 Data Recorder 6 ...
Страница 174: ...6 Data Recorder Documenting measurements 6 42 Edition 4 05 2014 DMS Go ...
Страница 175: ...DMS Go Edition 4 05 2014 7 1 Special functions of the DMS Go TC 7 ...
Страница 189: ...DMS Go Edition 4 05 2014 8 1 Maintenance and care 8 ...
Страница 195: ...DMS Go Edition 4 05 2014 9 1 Interfaces and Peripherals 9 ...
Страница 199: ...DMS Go Edition 4 05 2014 10 1 Appendix 10 ...
Страница 227: ...Recycling directives 10 Appendix DMS Go Edition 4 05 2014 10 29 10 9 Recycling directives Overview 2 3 3 ...
Страница 228: ...10 Appendix Recycling directives 10 30 Edition 4 05 2014 DMS Go 4 5 1 4 ...
Страница 230: ...10 Appendix Recycling directives 10 32 Edition 4 05 2014 DMS Go Materials to be disposed of separately 1 ...
Страница 231: ...Recycling directives 10 Appendix DMS Go Edition 4 05 2014 10 33 2 ...
Страница 233: ...Recycling directives 10 Appendix DMS Go Edition 4 05 2014 10 35 Other materials and components 2 3 3 ...
Страница 234: ...10 Appendix Recycling directives 10 36 Edition 4 05 2014 DMS Go 1 3 ...
Страница 235: ...Recycling directives 10 Appendix DMS Go Edition 4 05 2014 10 37 1 1 ...
Страница 236: ...10 Appendix Recycling directives 10 38 Edition 4 05 2014 DMS Go 5 ...
Страница 237: ...Recycling directives 10 Appendix DMS Go Edition 4 05 2014 10 39 4 7 ...
Страница 238: ...10 Appendix Recycling directives 10 40 Edition 4 05 2014 DMS Go 3 6 6 6 7 ...
Страница 242: ...10 Appendix Recycling directives 10 44 Edition 4 05 2014 DMS Go ...
Страница 243: ...DMS Go Edition 4 05 2014 11 1 Specifications 11 ...
Страница 252: ...11 10 Edition 4 05 2014 DMS Go ...
Страница 253: ...DMS Go Edition 4 05 2014 12 1 Index 12 ...