7 Special functions of the DMS Go+ TC
Measurement mode TopCOAT
7-10
Edition 4 (05/2014)
DMS Go+
Zeroing for TopCOAT
ATTENTION
The calibration described below is an abso-
lute prerequisite for the subsequent mea-
surements.
Checking reference block data
A special velocity reference block made of copper be-
longs to every DMS Go+ TC. The data of this reference
block are usually already stored in your DMS Go+ TC.
At first, check that these data (sound velocity and serial
number) correspond to those of the reference block. If
necessary, you can enter new data, e.g. if the original
reference block has been lost (see Chapter
The sound velocity (V) in inch/µs and m/s, and the serial
number (S/N) are engraved on two sides of the refer-
ence block.
Carrying out probe zero offset
ATTENTION
It is absolutely necessary that you use low-
viscosity couplant for the probe zero offset
(e.g. light machinery oil or glycerin).
Note
Before starting the calibration, set an initial
sound velocity between 5000 and 6000 m/s
in order to reduce the calibration time.
– In the function group
PROBE&CAL
, select the func-
tion
TG MODE
.
– Choose the option
TOP-COAT
.
– In the function group
PROBE&CAL
, select the func-
tion
VELOCITY
.
– Set a value between 5000 and 6000 m/s.
Содержание DMS Go+
Страница 1: ...DMS Go Technical Reference and Operating Manual ...
Страница 3: ...DMS Go Edition 4 05 2014 0 3 Function groups and functions ...
Страница 4: ...0 4 Edition 4 05 2014 DMS Go Function groups and functions cont d ...
Страница 8: ...0 8 Edition 4 05 2014 DMS Go ...
Страница 17: ...DMS Go Edition 4 05 2014 1 1 Introduction 1 ...
Страница 41: ...DMS Go Edition 4 05 2014 2 1 Standard package and accessories 2 ...
Страница 54: ...2 Standard package and accessories Recommended accessories 2 14 Edition 4 05 2014 DMS Go ...
Страница 55: ...DMS Go Edition 4 05 2014 3 1 Initial start up 3 ...
Страница 67: ...DMS Go Edition 4 05 2014 4 1 Principles of operation 4 ...
Страница 78: ...4 12 Edition 4 05 2014 DMS Go ...
Страница 79: ...DMS Go Edition 4 05 2014 5 1 Operation 5 ...
Страница 133: ...DMS Go Edition 4 05 2014 6 1 Data Recorder 6 ...
Страница 174: ...6 Data Recorder Documenting measurements 6 42 Edition 4 05 2014 DMS Go ...
Страница 175: ...DMS Go Edition 4 05 2014 7 1 Special functions of the DMS Go TC 7 ...
Страница 189: ...DMS Go Edition 4 05 2014 8 1 Maintenance and care 8 ...
Страница 195: ...DMS Go Edition 4 05 2014 9 1 Interfaces and Peripherals 9 ...
Страница 199: ...DMS Go Edition 4 05 2014 10 1 Appendix 10 ...
Страница 227: ...Recycling directives 10 Appendix DMS Go Edition 4 05 2014 10 29 10 9 Recycling directives Overview 2 3 3 ...
Страница 228: ...10 Appendix Recycling directives 10 30 Edition 4 05 2014 DMS Go 4 5 1 4 ...
Страница 230: ...10 Appendix Recycling directives 10 32 Edition 4 05 2014 DMS Go Materials to be disposed of separately 1 ...
Страница 231: ...Recycling directives 10 Appendix DMS Go Edition 4 05 2014 10 33 2 ...
Страница 233: ...Recycling directives 10 Appendix DMS Go Edition 4 05 2014 10 35 Other materials and components 2 3 3 ...
Страница 234: ...10 Appendix Recycling directives 10 36 Edition 4 05 2014 DMS Go 1 3 ...
Страница 235: ...Recycling directives 10 Appendix DMS Go Edition 4 05 2014 10 37 1 1 ...
Страница 236: ...10 Appendix Recycling directives 10 38 Edition 4 05 2014 DMS Go 5 ...
Страница 237: ...Recycling directives 10 Appendix DMS Go Edition 4 05 2014 10 39 4 7 ...
Страница 238: ...10 Appendix Recycling directives 10 40 Edition 4 05 2014 DMS Go 3 6 6 6 7 ...
Страница 242: ...10 Appendix Recycling directives 10 44 Edition 4 05 2014 DMS Go ...
Страница 243: ...DMS Go Edition 4 05 2014 11 1 Specifications 11 ...
Страница 252: ...11 10 Edition 4 05 2014 DMS Go ...
Страница 253: ...DMS Go Edition 4 05 2014 12 1 Index 12 ...