General System Requirements
1.3 Delivery Requirements
1.3.4 Crated and Uncrated Weights, Measurements and Clearance
Discovery NM/CT 670 – Standard Integration Instructions
5483063-1EN, Rev.2, ©2015 GE Healthcare
1-10
1.3.4 Crated and Uncrated Weights, Measurements and Clearance
The following tables provide you with crate and component measurements, weights and other
data, in order to assist you in planning conveyance routes and storage areas. The order of the
components in the list constitutes the recommended order of conveyance and delivery to the scan
room for installation. The information is provided in metric and imperial formats, as follows:
Table 1-1, D670 Components and Clearance — Metric, p.1-10
CT-specific Components and Clearance — Metric, p.1-12
Table 1-3, D670 Components and Clearance — Imperial, p.1-13
CT-specific Components and Clearance — Imperial, p.1-15
Table 1-1: D670 Components and Clearance — Metric
Crated
Uncrated
Component name
Crate size
(Dollies not attached)
(cm) (Height × Width ×
Depth/Length)
Weight
(kg)
Minimal dimensions
(cm)
*
Weight
(kg)
Door
width
Corridor /
elevator
width
Corridor /
elevator
length
Width of
corridors with
90° turns
†
Height
Pre-installation kit
‡
75×40×175
15
any
any
any
any
any
15
NM gantry with detectors
220×120×225
2230
140
140
222
250
200
2055
(with dollies)
Table
140×90×300
737
100
100
283
250
any
565
NM acquisition station
80×60×60
30
any
any
any
any
any
20
Peripherals and accessories
115×100×150
50
any
any
any
any
any
50
Collimators on carts
170×90×115
370
(for heaviest set)
55
55
100
112
150
330
(heaviest set)
CT gantry
231×137×200
2150
115
115
281
250
200
2010
(with dollies)
NM/CT covers
207×105×197
90
any
any
any
any
any
25