General System Requirements
1.3 Delivery Requirements
1.3.4 Crated and Uncrated Weights, Measurements and Clearance
Discovery NM/CT 670 – Standard Integration Instructions
5483063-1EN, Rev.2, ©2015 GE Healthcare
1-12
Table 1-2: D670 Additional CT-specific Components and Clearance — Metric
Crated
Uncrated
Component
name
Crate size
(Dollies not attached)
(cm) (Height × Width ×
Depth/Length)
Weight
(kg)
Minimal dimensions
(cm)
*
Weight
(kg)
Door
width
Corridor /
elevator
width
Corridor /
elevator
length
Width of
corridors with
90
°
turns
†
Height
CT PDU
123×90×70
D670/OPT
460
D670/BSE
440
100
100
100
100
any
D670/OPT
370
D670/BSE
350
CT UPS
82×31×81
any
any
any
any
any
160
CT console (NIO)
65×47×83
115
any
any
any
any
any
88
CT accessories
115×100×150
50
100
110
150
180
any
50
CT sub-system
covers
190×100×170
537
80
80
100
128
199
30
MDP (Main
Disconnect Panel)
‡
**
100×70×50
60
any
any
any
any
any
45
Optional Items
Contrast Injector
May vary but not
more than
80×80×80
May vary but not
more than 15
any
any
any
any
any
<13
AW W/S
*
The minimum door width required in order to bring the system components into the scan room also depends on the width of the corridor leading to
the room. When planning or measuring the width of the scan room door, use the graphs provided in
Figure 1-2: D670 Relative Required Width for
Corridor and Scan Room Door, p.1-17
in order to verify that the measurements comply with the requirements.
†
The corridor width required in order to move the system components from the unloading area to the scan room depends on the
angles of turns on the corridor. See
Figure 1-3: D670 Required Corridor Width for 90° Turns,p.1-18
for the required width when the angle is 90°.
‡
A1 or other; if ordered from GE, delivered in advance as part of the site preparation and must be installed prior to system installation
** In some cases, an alternative MDP is supplied, in order to comply with local electricity standards, see