Programming Site Data
1-33
February 2009
Entering a Calibration
Factor
The final option in the Program Menu, Correction Factor, allows you
to enter and modify the Calibration Factor. To enter the Correction
Factors option, scroll to the Correction Factors entry on the Program
Menu and press
[ENTER]
. The screen appears similar to Figure 1-32
below. To access the Calibration Factor, press the [
T
] key. Refer to
the
Program Menu Map,
Figure 1-34 on page 1-37.
Figure 1-32: Calibration Factor Window
1.
To enable or disable the Calibration Factor, scroll down to the next
line and select
On
or
Off.
a.
Use the [
W
] and [
X
] keys to scroll to the appropriate option.
b.
Press
[ENTER]
to confirm your selection.
2.
If
On
is selected for the Calibration Factor, scroll down to choose
a
Single
K-Factor or a
Table
of K-Factors.
a.
Use the [
W
] and [
X
] keys to scroll to the appropriate radio
button.
b.
Press
[ENTER]
to confirm your selection.
Inputs
1
Data Source
Meter K-Factor
OK
Cancel
On
Single
Off
Table
Calibration Factor
K Factor
NO UNIT
Edit Table
Correction Factors
Calibration Factor
Содержание CTF878
Страница 1: ...GE Sensing Model CTF878 Clamp On Tag Flowmeter Programming Manual ...
Страница 2: ...GE Sensing Model CTF878 Clamp On Tag Flowmeter Programming Manual 910 254PA4 February 2009 ...
Страница 7: ...Chapter 1 ...
Страница 47: ...Chapter 2 ...
Страница 59: ...Chapter 3 ...
Страница 60: ...Creating and Managing Sites Introduction 3 1 Site Manager 3 2 ...
Страница 73: ...Chapter 4 ...
Страница 89: ...Chapter 5 ...