February 2009
1-22
Programming Site Data
Entering Analog Outputs
(cont.)
Figure 1-23: Data Source Selection Window
4.
The next prompt requests the span (full scale) value for the analog
output. This value represents the 20 mA output (in flow units).
a.
Press
[ENTER]
to open the text box.
b.
Use the numeric keys to enter the desired value.
c.
Press
[ENTER]
to confirm your selection.
5.
The last prompt, On Error, asks you to select how the CTF878 will
handle the analog outputs in the event of a fault condition. The
meter offers three alternatives:
•
Hold Last Value (hold the last good reading)
•
Force Low (force the reading to 0 or 4 mA)
•
Force High (force the reading to 20 mA).
a.
Press
[ENTER]
to open the drop-down menu.
b.
Scroll to the desired selection.
c.
Press
[ENTER]
to confirm your selection.
You have completed entering data in the Analog Output option.
•
To confirm the entries and return to Operate mode, press
[F3]
(OK).
The CTF878 returns to Operate Mode.
•
To leave the window without confirming the entries, press
[F2]
(Cancel) or the
[ESC]
key. The CTF878 returns to Operate Mode.
OK
Cancel
No Unit
AuxIn
Pressure
Volume
Fwd Totalizer
Rev Totalizer
Temperature
m/s
DIAG
StdFwd Totalizer
Select Measurement
Velocity
Содержание CTF878
Страница 1: ...GE Sensing Model CTF878 Clamp On Tag Flowmeter Programming Manual ...
Страница 2: ...GE Sensing Model CTF878 Clamp On Tag Flowmeter Programming Manual 910 254PA4 February 2009 ...
Страница 7: ...Chapter 1 ...
Страница 47: ...Chapter 2 ...
Страница 59: ...Chapter 3 ...
Страница 60: ...Creating and Managing Sites Introduction 3 1 Site Manager 3 2 ...
Страница 73: ...Chapter 4 ...
Страница 89: ...Chapter 5 ...