Revision C
250cx Series Maternal/Fetal Monitor
6-19
2036947-001
Parts List, Drawings, and Replacement: Assembly/Disassembly of FRUs
4.
Disconnect 8-conductor Inverter cable from DSP board connector J1.
5.
Disconnect 20-conductor DSP-to-LCD Decoder ribbon cable from DSP board
connector J2.
6.
Disconnect 10-conductor DSP-to-Keypad ribbon cable from DSP board
connector J5.
7.
Remove one flat-head screw from each side of the monitor fastening the front
bezel to chassis.
8.
Open Recorder door, as it will interfere with front bezel removal.
9.
Tilt the front bezel forward, away from the chassis.
10. Disconnect 20-conductor Recorder ribbon cable from Recorder board connector
J2.
11. Disconnect power supply cable from J1 connector on Recorder board.
12. Disconnect Recorder EMC ground wire Fast-on connector from chassis tab.
13. Remove four screws fastening Recorder assembly mounting bracket to chassis.
14. Remove Recorder assembly and mounting bracket. Do not snag bracket on
hoses or cables while removing.
15. Flip over the Recorder assembly. Remove three screws fastening Recorder
assembly to mounting bracket.
Installing new Recorder Assembly:
1. Fasten new Recorder assembly to mounting bracket with three
screws.
2.
Fasten Recorder assembly and mounting bracket to chassis with four screws.
Do not fully tighten four screws, as you will need to adjust later in the
procedure.
3.
Connect Recorder EMC ground wire Fast-on connector to chassis tab.
4.
Connect power supply cable to J1 connector on Recorder board.
5.
Connect 20-conductor Recorder ribbon cable to Recorder board connector J2.
6.
Open Recorder door, as it will interfere with front bezel attachment.
7.
Tilt the front bezel back onto chassis. Ensure that two front bezel fiducial pins
are aligned with Recorder slots. Ensure that three tabs on the bottom of front
bezel are aligned with chassis slots. (See photograph below.)
8.
Refasten screws to each side of chassis.
9.
Close Recorder door.
10. Align the Recorder so that it is equidistant from both sides of the printer
opening in the front bezel and the Recorder door is flush with the outside of the
front bezel.
11. Check that door will open and close without interference.
12. Connect 10-conductor DSP-to-Keypad ribbon cable to DSP board connector J5.
13. Connect 20-conductor DSP-to-LCD Decoder ribbon cable to DSP board
connector J2.
Содержание Corometrics 250cx Series
Страница 2: ......
Страница 6: ...CE CE ii 0086 ...
Страница 14: ...viii 250cx Series Maternal Fetal Monitor Revision C 2036947 001 ...
Страница 16: ...1 2 250cx Series Maternal Fetal Monitor Revision C 2036947 001 For your notes ...
Страница 28: ...1 14 250cx Series Maternal Fetal Monitor Revision C 2036947 001 Introduction Service Requirements ...
Страница 29: ...Revision C 250cx Series Maternal Fetal Monitor 2 1 2036947 001 2 Equipment Overview ...
Страница 30: ...2 2 250cx Series Maternal Fetal Monitor Revision C 2036947 001 For your notes ...
Страница 54: ...2 26 250cx Series Maternal Fetal Monitor Revision C 2036947 001 Equipment Overview Theory of Operation ...
Страница 55: ...3 1 250cx Series Maternal Fetal Monitor Revision C 2036947 001 3 Installation ...
Страница 56: ...3 2 250cx Series Maternal Fetal Monitor Revision C 2036947 001 For your notes ...
Страница 81: ...Revision C 250cx Series Maternal Fetal Monitor 4 1 2036947 001 4 Maintenance ...
Страница 82: ...4 2 250cx Series Maternal Fetal Monitor Revision C 2036947 001 For your notes ...
Страница 140: ...4 60 250cx Series Maternal Fetal Monitor Revision C 2036947 001 Maintenance Preventative Maintenance Inspection Report ...
Страница 141: ...Revision C 250cx Series Maternal Fetal Monitor 5 1 2036947 001 5 Troubleshooting ...
Страница 142: ...5 2 250cx Series Maternal Fetal Monitor Revision C 2036947 001 For your notes ...
Страница 192: ...5 52 250cx Series Maternal Fetal Monitor Revision C 2036947 001 Troubleshooting General Troubleshooting ...
Страница 193: ...Revision C 250cx Series Maternal Fetal Monitor 6 1 2036947 001 6 Parts List Drawings and Replacement ...
Страница 194: ...6 2 250cx Series Maternal Fetal Monitor Revision C 2036947 001 For your notes ...
Страница 226: ...A 2 250cx Series Maternal Fetal Monitor Revision C 2036947 001 Technical Specifications For your notes ...
Страница 236: ...A 12 250cx Series Maternal Fetal Monitor Revision C 2036947 001 Technical Specifications Strip Chart Recorder ...
Страница 237: ...Revision C 250cx Series Maternal Fetal Monitor B 1 2036947 001 B Alarms Summary ...
Страница 238: ...B 2 250cx Series Maternal Fetal Monitor Revision C 2036947 001 Alarms Summary For your notes ...
Страница 241: ...Revision C 250cx Series Maternal Fetal Monitor C 1 2036947 001 C Electromagnetic Compatibility ...
Страница 242: ...C 2 250cx Series Maternal Fetal Monitor Revision C 2036947 001 Electromagnetic Compatibility For your notes ...
Страница 250: ...C 10 250cx Series Maternal Fetal Monitor Revision C 2036947 001 This page intentionally left blank ...
Страница 251: ......