250cx Series Maternal/Fetal Monitor
Revision C
2036947-001
Troubleshooting: System Troubleshooting
PRINTHEAD
Misalignment causing drag
DSP BOARD
Front panel keypad routed through DSP board
FAN
Secondary failure due to overheating only
VOLUME CONTROLS
UI KEYPAD BOARD
Would affect all buttons on front panel
MAIN BOARD
Communicates with keypad
VOLUME KEYPAD
Affects volume and alarm cancel
UI TO VOLUME BOARD CABLE
Affects volume and alarm cancel
DSP TO UI KEYPAD CABLE
Affects all front panel functions
SYSTEM POWER SUPPLY
DSP BOARD
Front panel keypad routed through DSP board
FAN
Secondary failure due to overheating only
RECORDER BUTTON: RECORDER ON
RECORDER BOARD TO MAIN
BOARD CABLE
UI KEYPAD BOARD
Would affect all buttons on front panel
RECORDER BOARD
Drives the stepper motor
DSP TO UI KEYPAD CABLE
Affects all front panel functions
RECORDER MOTOR
Could have open winding or bad wire crimp in
connector
DOOR SWITCH
Will not print if main boards thinks door is open
MAIN BOARD
Communicates with keypad
SYSTEM POWER SUPPLY
DSP BOARD
Front panel keypad routed through DSP board
FAN
Secondary failure due to overheating only
UI BUTTON: UA REF
DSP TO UI KEYPAD CABLE
Affects all front panel functions
UI KEYPAD BOARD
Would affect all buttons on front panel
SYSTEM POWER SUPPLY
MAIN BOARD
Communicates with keypad
DSP BOARD
Front panel keypad routed through DSP board
250cx Series Service & Troubleshooting
PROBLEM
SUBASSEMBLY
COMMENTS
Содержание Corometrics 250cx Series
Страница 2: ......
Страница 6: ...CE CE ii 0086 ...
Страница 14: ...viii 250cx Series Maternal Fetal Monitor Revision C 2036947 001 ...
Страница 16: ...1 2 250cx Series Maternal Fetal Monitor Revision C 2036947 001 For your notes ...
Страница 28: ...1 14 250cx Series Maternal Fetal Monitor Revision C 2036947 001 Introduction Service Requirements ...
Страница 29: ...Revision C 250cx Series Maternal Fetal Monitor 2 1 2036947 001 2 Equipment Overview ...
Страница 30: ...2 2 250cx Series Maternal Fetal Monitor Revision C 2036947 001 For your notes ...
Страница 54: ...2 26 250cx Series Maternal Fetal Monitor Revision C 2036947 001 Equipment Overview Theory of Operation ...
Страница 55: ...3 1 250cx Series Maternal Fetal Monitor Revision C 2036947 001 3 Installation ...
Страница 56: ...3 2 250cx Series Maternal Fetal Monitor Revision C 2036947 001 For your notes ...
Страница 81: ...Revision C 250cx Series Maternal Fetal Monitor 4 1 2036947 001 4 Maintenance ...
Страница 82: ...4 2 250cx Series Maternal Fetal Monitor Revision C 2036947 001 For your notes ...
Страница 140: ...4 60 250cx Series Maternal Fetal Monitor Revision C 2036947 001 Maintenance Preventative Maintenance Inspection Report ...
Страница 141: ...Revision C 250cx Series Maternal Fetal Monitor 5 1 2036947 001 5 Troubleshooting ...
Страница 142: ...5 2 250cx Series Maternal Fetal Monitor Revision C 2036947 001 For your notes ...
Страница 192: ...5 52 250cx Series Maternal Fetal Monitor Revision C 2036947 001 Troubleshooting General Troubleshooting ...
Страница 193: ...Revision C 250cx Series Maternal Fetal Monitor 6 1 2036947 001 6 Parts List Drawings and Replacement ...
Страница 194: ...6 2 250cx Series Maternal Fetal Monitor Revision C 2036947 001 For your notes ...
Страница 226: ...A 2 250cx Series Maternal Fetal Monitor Revision C 2036947 001 Technical Specifications For your notes ...
Страница 236: ...A 12 250cx Series Maternal Fetal Monitor Revision C 2036947 001 Technical Specifications Strip Chart Recorder ...
Страница 237: ...Revision C 250cx Series Maternal Fetal Monitor B 1 2036947 001 B Alarms Summary ...
Страница 238: ...B 2 250cx Series Maternal Fetal Monitor Revision C 2036947 001 Alarms Summary For your notes ...
Страница 241: ...Revision C 250cx Series Maternal Fetal Monitor C 1 2036947 001 C Electromagnetic Compatibility ...
Страница 242: ...C 2 250cx Series Maternal Fetal Monitor Revision C 2036947 001 Electromagnetic Compatibility For your notes ...
Страница 250: ...C 10 250cx Series Maternal Fetal Monitor Revision C 2036947 001 This page intentionally left blank ...
Страница 251: ......