5011A EN 20020315
Re-erasing image plates with status ‘to be erased’
To re-erase an image plate which you have given the status ‘to be erased’ via
the ID station:
1
Check that the Centricity™ SP1001 is ready for operation:
• the Centricity™ SP1001 must display the operator main screen with ‘Ready’
status, e.g.:
• the status indicator at the top of the Centricity™ SP1001 must be green and be lit
constantly.
2
Insert the cassette into the cassette
slot.
The Centricity™ SP1001 will automatically erase the image plate. The display will
read:
When the Centricity™ SP1001 has erased the image plate, it displays the operator
main screen.
3
Remove the cassette from the cassette slot.