50
5011A EN 20020315
Troubleshooting checklist
A survey of possible problems is listed below. If corrective actions are
straightforward, they are given below. The more elaborate troubleshooting
procedures are explained in detail in
Chapter 4, ‘Preventive maintenance and
of the Centricity™ SP1001 Reference manual
General errors
Errors during operation
Error
Action
The Centricity™ SP1001
does not
start up.
Refer to
on
page
of the Reference manual.
Status field:
Error field:
ERROR
‘SERVICE XXXXX
’
Contact your local service organization.
Status field:
Error field:
ERROR
‘CODE XXXXX
’
MESSAGE 1
Message 2
Action
POWER SUPPLY OUT
OF TOLERANCE
1. Check setting of
voltage selector
switch on back
panel.
2. Check fuses of
the machine
3. Check supply
voltage.
Refer to
on page
of the
Reference manual.
Set-up display
STATUS
FUNCTION MODE
1ST MESSAGE ARRAY
2nd Message Array
Patient_Last_Name Sub_Exam
Patient_Last_Name Sub_Exam
Patient_Last_Name Sub_Exam
Station Name
ERROR
5011 A UM EN CR SP1001.fm Page 50 Friday, March 15, 2002 4:11 PM
Содержание Centricity SP1001
Страница 1: ...Centricity SP1001 User manual 5011 A UM EN CR SP1001 fm Page 1 Friday March 15 2002 4 11 PM ...
Страница 4: ...4 5011A EN 20020315 5011 A UM EN CR SP1001 fm Page 4 Friday March 15 2002 4 11 PM ...
Страница 46: ...46 5011A EN 20020315 5011 A UM EN CR SP1001 fm Page 46 Friday March 15 2002 4 11 PM ...
Страница 55: ...Equipment information sheet Appendix A 5011 A UM EN CR SP1001 fm Page 55 Friday March 15 2002 4 11 PM ...
Страница 59: ...59 5011A EN 20020315 5011 A UM EN CR SP1001 fm Page 59 Friday March 15 2002 4 11 PM ...