15
5011A EN 20020315
The display
The Centricity™ SP1001 control panel has a backlit LCD display with 8 lines of
40 characters each. Its lay-out depends on the operating mode.
◆
In operator mode, the display has dedicated areas for specific information:
1
Set-up of image processing station:
• [blank]: Default image processing station selected.
• Off line: Transmission to all image processing stations disabled.
• [process.station] not ready: Image processing station not available.
• [process.station] rerouted: Images rerouted to other image processing
station.
2
Type of message
3
Extra comment or action to take
4
System status:
• READY: The Centricity™ SP1001 is ready for operation.
• BUSY: The Centricity™ SP1001 is treating an image plate.
• ERROR: An error has occurred. Refer to
on
page
• LOCKED: id.
• WARNING: id.
3
1
Set-up
STATUS
OPERATION MODE
1ST MESSAGE
2nd Message
Patient_Last_Name Sub_Exam
Patient_Last_Name Sub_Exam
Patient_Last_Name Sub_Exam
Station Name
ERROR
5
6
8.1
8.2
8.3
1
2
4
7
5011 A UM EN CR SP1001.fm Page 15 Friday, March 15, 2002 4:11 PM
Содержание Centricity SP1001
Страница 1: ...Centricity SP1001 User manual 5011 A UM EN CR SP1001 fm Page 1 Friday March 15 2002 4 11 PM ...
Страница 4: ...4 5011A EN 20020315 5011 A UM EN CR SP1001 fm Page 4 Friday March 15 2002 4 11 PM ...
Страница 46: ...46 5011A EN 20020315 5011 A UM EN CR SP1001 fm Page 46 Friday March 15 2002 4 11 PM ...
Страница 55: ...Equipment information sheet Appendix A 5011 A UM EN CR SP1001 fm Page 55 Friday March 15 2002 4 11 PM ...
Страница 59: ...59 5011A EN 20020315 5011 A UM EN CR SP1001 fm Page 59 Friday March 15 2002 4 11 PM ...