34
5011A EN 20020315
3
Insert the cassette containing the emergency image plate into the cassette slot
as shown below.
Make sure to insert the cassette with the hinge [1] at the top and the locking
mechanism [2] at the bottom.
The image plate will be digitized using the speed class, i.e. the sensitivity,
corresponding to the emergency button as defined during configuration.
When the Centricity™ SP1001 has treated the emergency image plate, it displays
the operator main screen. The
digital image data
are transmitted to the image
processing station accompanied by default ID data. The image processing station will
give the emergency image priority over the other images in the image processing
queue.
❖
If you do not enter a cassette within 15 seconds after pressing the
emergency button, the Centricity™ SP1001 will quit the emergency button
function and return to the operator main screen.
❖
To change the speed class corresponding to the emergency button, contact
your local service organization.
4
Remove the cassette from the cassette slot.
2
1
5011 A UM EN CR SP1001.fm Page 34 Friday, March 15, 2002 4:11 PM
Содержание Centricity SP1001
Страница 1: ...Centricity SP1001 User manual 5011 A UM EN CR SP1001 fm Page 1 Friday March 15 2002 4 11 PM ...
Страница 4: ...4 5011A EN 20020315 5011 A UM EN CR SP1001 fm Page 4 Friday March 15 2002 4 11 PM ...
Страница 46: ...46 5011A EN 20020315 5011 A UM EN CR SP1001 fm Page 46 Friday March 15 2002 4 11 PM ...
Страница 55: ...Equipment information sheet Appendix A 5011 A UM EN CR SP1001 fm Page 55 Friday March 15 2002 4 11 PM ...
Страница 59: ...59 5011A EN 20020315 5011 A UM EN CR SP1001 fm Page 59 Friday March 15 2002 4 11 PM ...