EX
WARNING
EXPLOSION HAZARD! Always use non-flammable liquids, for
example 20% ethanol, to pack the column, to flush solvents out of
the column, and to fill an empty column before filling it with sol-
vents. Otherwise, air might enter the flow path and create an
explosive atmosphere when being mixed with vaporized solvent.
The user is responsible for preparing procedures to make sure that
only nonflammable liquids, for example 20% ethanol, are used for
flushing the column or for filling an empty column before filling it
with solvents.
EX
WARNING
EXPLOSION HAZARD! When preparing organic solvent based
buffers in an empty container, always first fill the container with
inert gas. Work quickly with open containers. Otherwise, oxygen
and organic solvents might mix and create a potentially explosive
atmosphere.
EX
WARNING
Before maintenance, or if the column is to be moved to a non EX-
rated area, make sure that any flammable solvents have been
flushed out from the column.
Personal protection
WARNING
Always use appropriate personal protective equipment during
operation and maintenance of BioProcess LPLC or MPLC columns.
20
LPLC and MPLC Columns Operating Instructions 28-9638-66 AC
2 Safety instructions
2.1 Safety precautions