Action
Step
Remove the bolts and fittings holding the distributor in place on the lower
surface of the bottom end plate. These bolts and distributor fittings are
easily removed using a Hex/Allen key.
4
A
B
C
Description
Part
Liquid distributor.
A
Liquid distributor O-ring.
B
Mount for one of the fittings used to bolt the liquid distributor in
place on the bottom end plate.
C
Remove the liquid distributor by applying compressed air via the bottom
mobile phase inlet. The compressed air will cause the liquid distributor to
rise upwards.
5
Remove the O-ring from the liquid distributor and clean the surfaces of the
liquid distributor with 100% ethanol.
6
Place a new O-ring on the liquid distributor and replace the distributor fittings
in their mounts.
7
Lower the liquid distributor onto the bottom end plate and bolt it into position.
8
Mount a new O-ring into the groove on the outer circumference of the bottom
end plate.
9
Set a new bed support O-ring in place in the upper groove of the bottom
end plate.
10
LPLC and MPLC Columns Operating Instructions 28-9638-66 AC
83
6 Maintenance