3300/25 Dual Accelerometer Input Monitor Operation Manual
60
Channel B Integrator/Gain Stage Location
CHANNEL B
Main Board
Filter Board
INTEGRATOR/GAIN STAGE
LOCATION
INSTALL REMOVE
INSTALL
REMOVE
*
Integrator/Gain Stage with
NO
Filtering
W5B,3B W3A,3C
W41,60
W61,85
W86
**
Filter
after
the
Integrator/Gain Stage
W3B
W3A,3C
W5B
W41,60
W61,85
W86
Filter
before
the
Integrator/Gain Stage
W3A,3C
W5B
W3B W61
W41,60
W85,86
High Pass Filter
before
the
Integrator/Gain Stage
and Low Pass Filter
after
the
Integrator/Gain Stage
W3A,3C W3B,5B
W58,60
W85,86
W20,21
W41,59
W61
*
Denotes standard configuration.
**
When filtering without integration, set the filter
after
the Integrator/Gain Stage.
NOTE: If you have chosen the
NO Filtering
option, then you have completed your filter
Configuration.
When measuring velocity, low frequency signals can cause the monitor to saturate.
Содержание Bently Nevada 3300/25
Страница 11: ...Section 2 Monitor Functions 5 ...
Страница 51: ...Section 16 Self Test 45 When you press the RESET switch on the System Monitor the OK LEDs will stop flashing ...
Страница 93: ...Section 20 Appendix C Filter Board Options PWA 148921 01 87 20 2 Filter Board Jumper Locations ...
Страница 105: ...Section 20 Appendix C Filter Board Options PWA 148921 01 99 Page is blank ...