Section 20 - Appendix C–Filter Board Options PWA 148921-01
133
20.13 Test Filter Option (Alternate Procedure)
This procedure should only be used when your Full Scale Range Option is 05, 06, 15, 16,
or 17 (see
Section 8.2
) and the full scale amplitude calculated in step 5b of the main
procedure (
Section 20.12
) is greater than 3.5V or too small to set up accurately on your
function generator.
Note: Both channels A and B must be calibrated before performing this test (see
Section
15
). Test equipment should be set up as shown in step two of
Section 20.12
. If barriers
are used, test with barriers in place.
1.
Configure the main board as follows to change the Integrator Stage to a Gain
Stage:
Channel A:
Install
W6B,C
Remove
W6A
Channel B:
Install
W3D,E
Remove
W3F
2.
Adjust the frequency of the function generator to the -3dB frequency of the filter.
For High Pass Filters, the -3dB frequency is the corner frequency chosen in
Section 20.7 or 20.8
. For Low Pass Filters, the -3dB frequency is the corner
frequency chosen in
Section 20.10
. Set the amplitude of the function generator
for a full scale sine wave signal with a -7.5 Vdc offset:
3.
Verify that the front panel bargraph for the channel being tested reads 65% to
75% of full scale range.
4.
If you are testing a Band Pass Filter, repeat steps two and three at the second -
3dB frequency.
Full Scale Input Voltage
Full Scale Option
Vpk Vrms
05
0-1
in/s
06
0-2
in/s
15
0-25
mm/s
16
0-50
mm/s
17
0-100
mm/s
0.500
1.00
0.492
0.984
1.969
0.354
0.707
0.348
0.696
1.392
Содержание Bently Nevada 3300/25
Страница 11: ...Section 2 Monitor Functions 5 ...
Страница 51: ...Section 16 Self Test 45 When you press the RESET switch on the System Monitor the OK LEDs will stop flashing ...
Страница 93: ...Section 20 Appendix C Filter Board Options PWA 148921 01 87 20 2 Filter Board Jumper Locations ...
Страница 105: ...Section 20 Appendix C Filter Board Options PWA 148921 01 99 Page is blank ...