Chapter 4. Programming Guide
20 | 8000/8100/8200 Series Instruction Manual-English
© 2010 Baker Hughes, a GE company – All rights reserved.
Command:
<SPACE>Q,<Speed setting><CR>
Reply:
n/a
Query:
<SPACE>Q,?<CR>
Reply:
<Measurement Speed><CR>
Query:
<SPACE>*Q,?<CR>
Reply:
Measurement Speed = <Measurement Speed><CR>
4.6.4 U - Set Units
This command sets the units of pressure measurement.
Command:
<SPACE>U,<Unit number><CR>
Reply:
n/a
Result:
Changes the pressure units of the transmitted reading.
Query:
<SPACE>U,?<CR>
Reply:
<Unit number><CR>
Command:
<SPACE>*U<CR>
Reply:
(Interactive)
Current units are <Units> (<unit number>)<CR>
Table 4-3: Measure Speed Setting
Q Command
Setting
Resonant
Sensor Cycles
Counted
Measurement Interval
a
(ms)
a. This interval varies between minimum and maximum.
G Command
Fixed Setting
Minimum
Maximum
0
64000
1600
2560
4.0
1
32000
800
1280
2.0
2
b
b. Default setting for the accuracy stated in the specification.
16000
400
640
1.0
3
8000
200
320
0.50
4
4000
100
160
0.25
5
2000
50
80
0.25
where: <Unit number>
is a number between 0 and 24 used to select the required units.