Appendix A. Fault Finding
34 | 8000/8100/8200 Series Instruction Manual-English
© 2010 Baker Hughes, a GE company – All rights reserved.
!019 Timed Out
Key not pressed in the specified time
period (usually 20 seconds) normally
happens with inactive commands.
Repeat command, entering response in
less than 20 seconds between each
character and send carriage return within
the specified time.
!020 No
Frequency
Change of measurement speed, see
Section 4.6.3 on page 19.
Wait for 2 seconds, if fault remains return
to GE for repair or replacement.
!021 Bad
Checksum
Error occurred in memory check sum.
Power cycle sensor, if fault remains return
to GE for repair or replacement.
!022 Bad
Message
Erroneous character detected in the
middle of a message.
Check and re-send message.
!023 Bad Cal
Pres
Divide by zero error occurred in
calibration calculation.
Check data and repeat calibration.
Table A-1: Fault Finding