B30 Patient Monitor
4-6
Alarm Volume
Adjusts the volume of the audio alarms. The range is from 1 (soft) to 10 (loud).
NOTE: Audible alarms cannot be totally silenced with the
Alarm Volume
function.
Alarm Light
Adjusts the brightness of the alarm light if the light has been enabled (default) through
Monitor Setup
-
Install/Service - Installation - Alarm Options - Alarm Light- YES
. The range
is from 5 to 100%. The light is on during the adjustment to help determine a suitable brightness.
Audio ON/OFF
Opens a menu to select
Silence Apnea
,
Silence ECG
,
Silence Apn & ECG
or
Silence All
.
NOTE: The silencing selections that is, all other selections than the
Activate Alarms
) are
available only if alarm silencing has been enabled by selecting
Monitor Setup
-
Install/
Service - Installation - Alarm Options - Show Audio ON/OFF - Yes
. By default, it is disabled. For
more details, see
”Enabling or disabling alarm silencing” page 4-14.
Activate Alarms
= Activates silenced alarms. This selection is always available for activating
alarms that have been permanently silenced using the Central.
Silence Apn
= Silences apnea and disconnection alarms as well as CO
2
, respiration rate.
Silence ECG
= Silences arrhythmia alarms and also HR limit alarms.
Silence Apn & ECG
= Silences both of above.
Silence ALL
= Silences permanently all alarms except FiO
2
<18%, Ppeak >70 cmH
2
O.
Remove Menu
Clears the menu selections from the display so that only 10 minute trends and limits are
displayed (push the ComWheel to return the selections on the screen).
Содержание b30
Страница 2: ......
Страница 8: ......
Страница 10: ...Table of contents ii...
Страница 12: ...Table of contents ii...
Страница 16: ...B30 Patient Monitor 4...
Страница 18: ...Table of contents ii...
Страница 26: ...Table of contents ii...
Страница 62: ...B30 Patient Monitor 2 36...
Страница 64: ...Table of contents ii...
Страница 72: ...Table of contents ii...
Страница 90: ...Table of contents ii...
Страница 106: ...Table of contents ii...
Страница 118: ...B30 Patient Monitor 6 12...
Страница 120: ...Table of contents ii...
Страница 124: ...B30 Patient Monitor 7 4...
Страница 126: ...Table of contents ii...
Страница 136: ...Table of contents ii...
Страница 144: ...B30 Patient Monitor 9 8...
Страница 146: ...Table of contents ii...
Страница 156: ...Table of contents ii...
Страница 176: ...Table of contents ii...
Страница 184: ...B30 Patient Monitor 12 8...
Страница 186: ...Table of contents ii...
Страница 192: ...Table of contents ii...
Страница 202: ...Table of contents ii...
Страница 212: ...B30 Patient Monitor 15 10...
Страница 214: ...Table of contents ii...
Страница 224: ...B30 Patient Monitor 16 10...
Страница 226: ...Table of contents ii...
Страница 236: ...B30 Patient Monitor 17 10...
Страница 241: ......