B30 Patient Monitor
15-8
Measurement limitations
Movement artifacts
Changing position, moving the head, moving the arms or shivering may result in movement
artifacts. Also the heart may cause noticeable movement and sometimes this may interfere
with the respiration measurement.
Electrical interference
Electrical devices, such as electrosurgery units and infrared heaters, that emit electromagnetic
disturbance may cause artifacts or disable the respiration measurement completely.
Intermittent mechanical ventilation
During spontaneous breathing the ventilator may at times support the patient’s ventilation
with an extra inspiration. If these ventilator inspirations are substantially larger than the
spontaneous breaths, the respiration calculation may mistakenly count only the inspirations
and expirations produced by the ventilator. You can correct this by adjusting the detection
limits manually:
1.
Press the
Others
key.
2.
Select
Resp Setup
.
3.
Select
Detection limit
and adjust the limits.
Turning off the measurement
If the impedance respiration measurement signal interferes with other measurements, such as
ECG, you can turn it off.
1.
Press the
Others
key.
2.
Select
Resp Setup
.
3.
Select
Measurement OFF
.
Others
Содержание b30
Страница 2: ......
Страница 8: ......
Страница 10: ...Table of contents ii...
Страница 12: ...Table of contents ii...
Страница 16: ...B30 Patient Monitor 4...
Страница 18: ...Table of contents ii...
Страница 26: ...Table of contents ii...
Страница 62: ...B30 Patient Monitor 2 36...
Страница 64: ...Table of contents ii...
Страница 72: ...Table of contents ii...
Страница 90: ...Table of contents ii...
Страница 106: ...Table of contents ii...
Страница 118: ...B30 Patient Monitor 6 12...
Страница 120: ...Table of contents ii...
Страница 124: ...B30 Patient Monitor 7 4...
Страница 126: ...Table of contents ii...
Страница 136: ...Table of contents ii...
Страница 144: ...B30 Patient Monitor 9 8...
Страница 146: ...Table of contents ii...
Страница 156: ...Table of contents ii...
Страница 176: ...Table of contents ii...
Страница 184: ...B30 Patient Monitor 12 8...
Страница 186: ...Table of contents ii...
Страница 192: ...Table of contents ii...
Страница 202: ...Table of contents ii...
Страница 212: ...B30 Patient Monitor 15 10...
Страница 214: ...Table of contents ii...
Страница 224: ...B30 Patient Monitor 16 10...
Страница 226: ...Table of contents ii...
Страница 236: ...B30 Patient Monitor 17 10...
Страница 241: ......