Impedance respiration
15-7
Correcting the respiration number
Normally, we recommend the use of the
AUTO
detection limit. However, in some specific cases
you may wish to adjust the limits manually.
−
When the respirations are weak, you can manually adjust the detection limits
(measurement sensitivity) closer to each other to ensure that all respirations are included
in the RR value. In this case, the dotted line represents the absolute detection limits.
−
When there are lots of artifacts, the grids can be adjusted further apart to separate
smaller artifacts from larger, true respiration peaks. The small peaks fall within the grids
and are not calculated, while the bigger peaks cross the grids and are calculated as true
respirations.
To ensure the correct respiration number, adjust the limits closer to each other:
1.
Press
Others
.
2.
Select
Resp Setup
-
Detection Limit
and adjust the limit.
The percentage is the ratio to the 1
Ω
reference bar which is 100%.
Others
Содержание b30
Страница 2: ......
Страница 8: ......
Страница 10: ...Table of contents ii...
Страница 12: ...Table of contents ii...
Страница 16: ...B30 Patient Monitor 4...
Страница 18: ...Table of contents ii...
Страница 26: ...Table of contents ii...
Страница 62: ...B30 Patient Monitor 2 36...
Страница 64: ...Table of contents ii...
Страница 72: ...Table of contents ii...
Страница 90: ...Table of contents ii...
Страница 106: ...Table of contents ii...
Страница 118: ...B30 Patient Monitor 6 12...
Страница 120: ...Table of contents ii...
Страница 124: ...B30 Patient Monitor 7 4...
Страница 126: ...Table of contents ii...
Страница 136: ...Table of contents ii...
Страница 144: ...B30 Patient Monitor 9 8...
Страница 146: ...Table of contents ii...
Страница 156: ...Table of contents ii...
Страница 176: ...Table of contents ii...
Страница 184: ...B30 Patient Monitor 12 8...
Страница 186: ...Table of contents ii...
Страница 192: ...Table of contents ii...
Страница 202: ...Table of contents ii...
Страница 212: ...B30 Patient Monitor 15 10...
Страница 214: ...Table of contents ii...
Страница 224: ...B30 Patient Monitor 16 10...
Страница 226: ...Table of contents ii...
Страница 236: ...B30 Patient Monitor 17 10...
Страница 241: ......