Modifications
Reserved
24
OPM_ITS_USM_M06_M20_0US_V011 (09/2007)
2000 VA Rack
Black Wire
From UPS
Red Wire
From UPS
Jumper Wire
(Re-Use)
FRONT
Jumper Wire
(Re-Use)
Jumper Wire
(Re-Use)
FRONT
Black Wire
From UPS
Red Wire
From UPS
Jumper Wire
(Re-Use)
Jumper Wire
(Re-Use)
1500 VA Rack
FRONT
Black Wire
From UPS
Red Wire
From UPS
Jumper Wire
(Re-Use)
600 & 1000 VA Rack
Step 7.
Disconnect Jumper wire(s) and remove batteries.
Step 8.
Install New Batteries using same battery type and layout as shown in diagrams above.
Step 9.
Connect Jumper wire(s) to the + and – of the new batteries as shown in diagrams above.
Step 10.
Before inserting the batteries into the unit: Connect the Black and Red Wire on the + terminal as shown in
diagrams above.
•
Slide the batteries into the UPS.
Step 11.
Install Metal Battery cover plate: Install metal plate with Phillips-head screws.
Step 12.
Re-Install Left front cover first by using the Phillips-head screws. Put the Right front cover on the Right side first;
press the left side into the UPS.
Step 13.
Battery replacement complete.