8.2
Chemical resistance
Introduction
The table below gives allowed exposure concentrations and times for various chemicals
that may be used in GE Healthcare BioProcess instruments in general.
Some of the chemicals listed may not be applicable for your instrument.
WARNING
Flammable liquids. ÄKTAprocess is not approved to handle
flammable liquids.
WARNING
Some of the chemicals used with ÄKTAprocess may be flammable
under certain conditions. Make sure to use chemicals only under
conditions where they are not flammable. Refer to local and/or
national classifications of flammable liquids.
List of allowed chemicals
Usage
Max acc. expos.
Max time / cycle
Concentration
Chemical
CIP
3000 h
3 h
25%
Acetic acid
UV cell test
Unlimited
1 h
10%
Acetone
CIP
1000 h
1 h at temp ≤ 60°C
pH 2 to 2.5
Citric acid
Storage
Unlimited
12 months
20%
Ethanol
CIP
3000 h
3 h
20%
Ethanol / acetic
acid
CIP
5000 h
5 h
6 M
Guanidine
hydrochloride
CIP
1000 h
1 h
0.1 M at pH=1
Hydrochloric
acid
ÄKTAprocess Operating Instructions 29-0252-49 AA
167
8 Reference information
8.2 Chemical resistance