6.4
Disassembly and assembly
Introduction
This section covers all disassembly and assembly procedures that the end user is allowed
to perform without support from GE Healthcare.
Precautions
WARNING
Only personnel authorized by GE Healthcare may open the cabinet
doors. There is high voltage inside the cabinet that can cause
human injury or death.
WARNING
The electric cabinet doors may only be opened when ÄKTAprocess
is taken out of operation and subject to LOCK OUT / TAG OUT.
WARNING
LOCK OUT / TAG OUT! Before any maintenance or decommission-
ing work is performed on ÄKTAprocess, make sure that:
•
it is empty and depressurized.
•
it is disconnected from process feed, electrical power and
pneumatic supply.
•
it is prevented from accidentally becoming re-energized during
maintenance.
•
it is clearly tagged as taken out of operation.
•
all process wetted areas are clean and decontaminated.
146
ÄKTAprocess Operating Instructions 29-0252-49 AA
6 Maintenance
6.4 Disassembly and assembly