5.1.3
Priming and leakage test
Introduction
Before a run is started the system should be primed and tested for possible leakage.
This section describes a procedure for priming and testing for leakage.
Important
NOTICE
All system parts must be correctly filled with liquid before any
functional tests are performed. A suitable liquid is a storage or
running buffer for the column.
NOTICE
During the procedure described below, check all valves and tubing
connections carefully for air or liquid leakage. If there is any evi-
dence of leakage, attend to the leak and then repeat the filling se-
quence.
Each pump must be tested separately if more than one is used (Optional Gra-
dient pump B and/or Sample pump in addition to the standard pump A).
Note:
Prepare for priming and leakage
test
Follow the instruction below to prepare ÄKTAprocess for priming and leakage test.
Action
Step
Check that the inlet and outlet tubing are correctly connected.
1
Check that the buffer tank(s) are filled with a sufficient amount of a suitable
liquid for the procedure.
2
Check that valves to the buffer tanks are open.
3
118
ÄKTAprocess Operating Instructions 29-0252-49 AA
5 Operation
5.1 Prepare the system
5.1.3 Priming and leakage test