RTfino/RT3200 Advantage–II/RT–x200 SERVICE MANUAL
GE MEDICAL SYSTEMS
P9030GX
INSTALLATION AND REPLACEMENT
3–26
REV 21
3–8–1 GAS SPRING REPLACEMENT (continued)
NECK
SUPPORT
BOLT
TIEWRAP
SCREWS
FAN
BOARDS
SIDE
BRACKET
REAR VIEW
ILLUSTRATION 3–17
18. Hold the pipe and raise it until the pin welded pipe reaches plastic of main board. Pull the pipe with a little twist.
19. Unscrew the nut fixing gas spring to pipe and remove the gas spring from the pipe. Hollow pipe without gas spring
looks as shown in Illustration 3–19.
HEXAGONAL NUT FIXING
GAS SPRING WITH PIPE.
BOTTOM VIEW – GAS SPRING
ILLUSTRATION 3–18
20. Put a new gas spring through the pipe and fix it with the nut.
21. Wind the cable four times around the gas spring and fix the cable to the gas spring with tape.