RTfino/RT3200 Advantage–II/RT–x200 SERVICE MANUAL
GE MEDICAL SYSTEMS
P9030GX
INSTALLATION AND REPLACEMENT
3–27
REV 21
3–8–1 GAS SPRING REPLACEMENT (continued)
PIPE
BELLOW
TAPE
GAS SPRING
HOLLOW PIPE
GAS SPRING ASSEMBLY
ILLUSTRATION 3–19
22. Put the assembly from step 21 back to the same position pulled from the main position. Be careful not to get the
cable in between bearing and pipe.
23. Remove the tape fixing cable in step 21 from the low part of the main body and fix the part of cable covered the
heat compression tube on neck support with tiewrap.
24. Insert the screw removed in step 17 to original position and fix the gas spring.
25. Fix two earth cables removed in step 15 with two screws.
26. Install the side bracket removed in step 14.
27. Install the signal cable and power supply cable removed in step 13 to the mother board.
28. Install the front cover, front plate, front panel and rear cover removed in step 12.
29. Install the monitor arm removed in step 11 to pipe.
30. Install the push bracket removed in step 8 with insert from the upper part to the monitor arm.
31. Install the clamp plate removed in step 9. Be careful as stud is easy to break.
32. Install the monitor arm cover removed in step 7 to the monitor arm.