RTfino/RT3200 Advantage–II/RT–x200 SERVICE MANUAL
GE MEDICAL SYSTEMS
P9030GX
INSTALLATION AND REPLACEMENT
3–6
REV 21
3–5 ASSEMBLING THE RTfino/RT3200 Advantage–II/RT–x200 SYSTEM
3–5–1 Installation
1.
Locate remove, and unpack the transducers located in the separate container.
2.
Secure any of transducers to either of the two receptacle as follows.
a. Ensure transducer twist lock lever points towards the 2 o’clock position.
b. Install transducer connector on the receptacle guide pin until it touches the receptacle mating surface.
c. Twist the transducer connector twist lock lever to the 6 o’clock position to lock it in place. Twist the lever to the 2
o’clock position to disconnect the transducer.
3.
It is not necessary to turn off to RTfino/RT3200 Advantage–II/RT–x200 system power to connect or disconnect a
transducer.
4.
Unpack the foot switch. Connect the foot switch to the Connector on the lower RTfino/RT3200 Advantage–II/RT–
x200 front panel.
5.
Connect the RTfino/RT3200 Advantage–II/RT–x200 power cable to a hospital grade power receptacle of the
proper voltage. Never use a three–to–two prong adapter; this defeats the safety ground.