Section 8-2 - DISASSEMBLY/RE-ASSEMBLY
8 - 7
8-2-2
OSD button
8-2-2-1
Tools
•
L wrench
•
Philips screw driver
8-2-2-2
Preparation
Shut down the system and switch off the main breaker
8-2-2-3
17" LCDMON OSD Key Assy Replace Procedure
1.) Remove the LCD Cable cover. Refer to the figure below.
2.) Unscrew 4 screws to remove the rear cover.Refer to the figure below.
Figure 8-9 Removing the LCD Cable cover
Figure 8-10 Removing the rear cover