7.) ARM FOLD COVER
Unscrew 1 screw (2329677, TAP M4X16) to remove the Arm fold cover. Refer to the figure below.
8.) Perform the following functional tests. If all are successful, include the debrief script provided below.
8-2-5-4
Mounting procedure
Install the each of the new part in the reverse order of removal procedure.
Figure 8-69 Removing the Arm fold cover
Table 8-6
Functional Tests
Service Manual
Section
Functional Test / Diagnostic Test
Debrief Script
“Service Manual, Direction
5245279, Rev 3+, Section 8-2-5. Equipment
passed all required tests and is ready for use. “